Sense Subsystem; Adjacent Channel Power Measurement - Agilent Technologies E4406A VSA Series Programmer's Manual

Transmitter tester
Hide thumbs Also See for E4406A VSA Series:
Table of Contents

Advertisement

Language Reference

SENSe Subsystem

SENSe Subsystem
Sets the instrument state parameters so that you can measure the
input signal.

Adjacent Channel Power Measurement

Commands for querying the adjacent channel power measurement
results and for setting to the default values are found in the
Group of Commands" on page
the parameters described in the following commands, are found under
the
key, after the
Meas Setup
the
key menu.
MEASURE
Adjacent Channel Power—Average Count
[:SENSe]:ACP:AVERage:COUNt <integer>
[:SENSe]:ACP:AVERage:COUNt?
Set the number of data acquisitions that will be averaged. After the
specified number of average counts, the average mode (termination
control) setting determines the average action.
Factory Preset
and *RST:
10, for cdma2000, W-CDMA mode
20, for Basic, cdmaOne, iDEN mode
Range:
1 to 10,000
Remarks:
Use INSTrument:SELect to set the mode.
Adjacent Channel Power—Averaging State
[:SENSe]:ACP:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:ACP:AVERage[:STATe]?
Turn average on or off.
Factory Preset
and *RST:
On
Off, for iDEN mode
Remarks:
Use INSTrument:SELect to set the mode.
210
187. The equivalent front panel keys for
measurement has been selected from
ACP
"MEASure
Chapter 5