Language Reference
SENSe Subsystem

RF Port Input Attenuation

[:SENSe]:POWer[:RF]:ATTenuation <rel_power>
[:SENSe]:POWer[:RF]:ATTenuation?
Set the RF input attenuator. This value is set at its auto value if input
attenuation is set to auto.
Factory Preset
and *RST:
0 dB
12.0 dB for iDEN
Range:
0 to 40 dB
Default Unit:
dB
Front Panel
Access:
Input, Input Atten

RF Port Power Range Maximum Total Power

[:SENSe]:POWer[:RF]:RANge[:UPPer] <power>
[:SENSe]:POWer[:RF]:RANge[:UPPer]?
Set the maximum expected total power level at the radio unit under
test. This value is ignored if RF port power range is set to auto.
External attenuation required above 30 dBm.
Factory Preset
−15.0 dBm
and *RST:
−100 to 80 dBm for GSM
Range:
−100 to 27.7 dBm for cdmaOne, iDEN
−200 to 50 dBm for NADC, PDC
−200 to 100 dBm for cdma2000, W-CDMA
Default Unit:
dBm
Remarks:
Global to the current mode. This is coupled to the RF
input attenuation
You must be in the Service, cdmaOne, GSM, NADC,
PDC, cdma2000, W-CDMA mode to use this command.
Use INSTrument:SELect to set the mode.
Front Panel
Access:
Input, Max Total Pwr (at UUT)