Cabling; Scan Mode; Ac Adapter; Manual Calibration Process - Fluke 1523 Technical Manual

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

7.8.2

Cabling

7.8.3

Scan Mode

7.8.4

AC Adapter

7.9 Manual Calibration Process

NOTE: 1. The 1523/24 is tested and calibrated using the tests outlined below in the
order indicated. Details on performing each of the tests can be found in the subsequent
section of this document.
2. For test specifications, durations, soak times, and other detailed information, refer
to the specification tables below.
Table 28 1523/24 Accuracy Test Settings and Specifications
Resistance
Range
L75_OHMS
LO_OHMS
MED_OHMS
Calibration of Your Reference Thermometer Readout
Reference
(Nominal) (Ω)
Soak (Sec)
0
20
25
20
75
20
75
20
100
20
200
20
400
20
200
20
400
20
10k
20
40k
20
Manual Calibration Process
Spec
Samples
(±Ω)
40
0.002
40
0.003
40
0.005
40
0.005
40
0.006
40
0.01
40
0.018
40
0.52
40
0.54
40
1.5
40
4.5
Std Dev
Spec (±Ω)
0.0005
0.00075
0.00125
0.00125
0.0015
0.0025
0.0045
0.13
0.135
0.375
1.125
81

Advertisement

Table of Contents
loading

This manual is also suitable for:

1524

Table of Contents