1523, 1524 Reference Thermometer
Manual Calibration Process
39. Turn off T1 by sending the following command.
40. Connect a 100 Ohm resistor to the calibration cable connected to the UUT on
41. Query the reading every 2 seconds for 40 readings using the following
42. Turn off Channel T2 by sending the serial command:
7.9.1.3.3
Medium Ohms Range
43. Set the UUT in the MED_OHMS cal range by sending the following commands
44. Connect a 200 ohm resistor to the calibration cable connected to the UUT.
45. Query the reading every 2 seconds for 40 readings using the following
46. Connect a 400 Ohm resistor to the calibration cable connected to the UUT.
47. Query the reading every 2 seconds for 40 readings using the following
48. Connect a 10K Ohm resistor to the calibration cable connected to the UUT.
49. Query the reading every 2 seconds for 40 readings using the following
50. Connect a 40K Ohm resistor to the calibration cable connected to the UUT.
51. Query the reading every 2 seconds for 40 readings using the following
Note: Steps 52 through 56 apply to the 1524 only. Proceed to step 57 if using a 1523.
86
Channel T2.
command:
through the serial port:
command:
command:
command:
command: