7.9.3

1523/24 As Left Test Data

1. Take as Left data by following the steps in the As Found Data section.
2. Calculate the average and the standard deviation of the measurements taken in
step 1, including the Channel T2 data. Record the results.
3. Compare the error with the specification in the Accuracy Test Setting and
Specification tables for each range to determine the pass/fail status. Record the
result.
Calibration of Your Reference Thermometer Readout
Manual Calibration Process
93