Fluke 1523 Technical Manual page 93

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

25. Connect a 25 Ohm resistor (28 ppm) to the calibration cable connected to the
UUT.
26. Query the reading every 2 seconds for 40 readings using the following
command:
27. Connect a 75 Ohm resistor to the calibration cable connected to the UUT.
28. Query the reading every 2 seconds for 40 readings using the following
command:
7.9.1.3.2
Lo Ohms Range
29. Set the UUT in the LO_OHMS cal range by sending the following commands
through the serial port:
30. Connect a 75 Ohm resistor to the calibration cable connected to the UUT.
31. Query the reading every 2 seconds for 40 readings using the following
command:
32. Connect a 100 Ohm resistor to the calibration cable connected to the UUT.
33. Query the reading every 2 seconds for 40 readings using the following
command:
34. Connect a 200 Ohm resistor to the calibration cable connected to the UUT.
35. Query the reading every 2 seconds for 40 readings using the following
command:
36. Connect a 400 Ohm resistor to the calibration cable connected to the UUT.
37. Query the reading every 2 seconds for 40 readings using the following
command:
Note: Steps 38 through 42 apply to the 1524 only. Proceed to step 43 if using a 1523.
38. (1524 ONLY) Set Channel T2 of the UUT in the LO_OHMS cal range by
sending the following commands through the serial port:
Calibration of Your Reference Thermometer Readout
Manual Calibration Process