1523/24 Accuracy Test (As Found) - Fluke 1523 Technical Manual

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

1523, 1524 Reference Thermometer
Manual Calibration Process
15. Set the UUT in the MV_OHMS cal range by sending the following commands
16. Query the Ref points by sending the following Command:
17. Query the ADJ values for the range by sending the following Command:
18. Record the as found alignment parameters.
7.9.1.3

1523/24 Accuracy Test (As Found)

19. Connect the calibration cable, described in the Preparation and Settings for
20. Connect an electrical short (less than 0.00054 ) to all 4 wires of the calibration
21. Set the UUT password enable by sending the following commands through the
22. Query the password enable to ensure communication is working and that it is
7.9.1.3.1
L75 Ohms Range
23. Set the UUT in the L75_OHMS cal range by sending the following commands
24. Query the reading every 2 seconds for 40 readings using the following
84
through the serial port:
Reference readout Calibration section, to Channel T1 of the UUT. Using a
terminal program set the terminal program communication settings to:
Baud
9600
Parity
None
Flow
Control None
Stop Bits
1
cable connected to the UUT.
serial port:
enabled before continuing. Do this by sending the following command:
through the serial port: