Fluke 1523 Technical Manual page 95

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

52. (1524 ONLY) Set Channel T2 of the UUT in the MED_OHMS cal range by
sending the following commands through the serial port:
53. Turn off T1 by sending the following command.
54. Connect a 10K Ohm resistor to the calibration cable connected to the UUT on
Channel T2.
55. Query the reading every 2 seconds for 40 readings using the following
command:
56. Turn off Channel T2 by sending the serial command:
7.9.1.3.4
High Ohms Range
57. Set the UUT in the HI_OHMS cal range by sending the following commands
through the serial port:
58. Connect a 40K Ohm resistor to the calibration cable connected to the UUT.
59. Query the reading every 2 seconds for 40 readings using the following
command:
60. Connect a 100K Ohm resistor to the calibration cable connected to the UUT.
61. Query the reading every 2 seconds for 40 readings using the following
command:
62. Connect a 300K Ohm resistor to the calibration cable connected to the UUT.
63. Query the reading every 2 seconds for 40 readings using the following
command:
64. Connect a 500K Ohm resistor to the calibration cable connected to the UUT.
65. Query the reading every 2 seconds for 40 readings using the following
command:
Calibration of Your Reference Thermometer Readout
Manual Calibration Process
87