Fluke 1523 Technical Manual page 98

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

1523, 1524 Reference Thermometer
Manual Calibration Process
Setting
Ref1, Ref2
Ref1, Ref2
Adj1 & Adj2
Adj1 & Adj2
Adj1 & Adj2
Adj1 & Adj2
Adj1 & Adj2
NOTE: All serial commands below must be sent in the order shown. Not doing so will
cause other calibration parameters to change values. This produces large errors in the
calibration.
1. Using a terminal program set the terminal program communication settings to:
2. Set the number of Ref points to 2 for each range by sending the following serial
3. Set all Reference points to their designated values (see Table 31 on previous
90
Cal Range
MED_OHMS
HI_OHMS
40000, 500000
MV
L75_OHMS
LO_OHMS
MED_OHMS
HI_OHMS
Baud
9600
Parity
None
Flow
Control None
Stop Bits
1
commands.
page, 1523/24 Alignment settings) for each range by sending the following
commands:
Value
Description
400, 40000
MED_OHMS Ref Calibration points
HI_OHMS Ref Calibration points
0,0
MV Ref Point Adjustment Values
0,0
L75_OHMS Ref Point Adjustment Values
0,0
LO_OHMS Ref Point Adjustment Values
0,0
MED_OHMS Ref Point Adjustment Values
0,0
HI_OHMS Ref Point Adjustment Values