Table 12 Temperature, Thermocouples External Reference Junction - Fluke 1523 Technical Manual

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

1523, 1524 Reference Thermometer
Specifications

Table 12 Temperature, Thermocouples External Reference Junction

Type
B
C
E
J
K
L
M
N
R
S
T
U
Resolution: 0.01 °C
Note 1: Accuracies are based on external reference junction
compensation.
20
Measure Accuracies
Range
(ITS-90)
600 to 800 °C
0.85 °C
800 to 1000 °C
0.68 °C
1000 to 1800 °C
0.57 °C
100 to 550 °C
0.32 °C
550 to 2300 °C
0.71 °C
-200 to 0 °C
0.18 °C
0 to 950 °C
0.09 °C
-200 to 0 °C
0.21 °C
0 to 1200 °C
0.12 °C
-200 to 0 °C
0.31 °C
0 to 1370 °C
0.15 °C
-200 to 0 °C
0.15 °C
0 to 900 °C
0.11 °C
-20 to 0 °C
0.14 °C
0 to 400 °C
0.14 °C
400 to 1400 °C
0.14 °C
-200 to 0 °C
0.48 °C
0 to 1300 °C
0.20 °C
-20 to 0 °C
1.06 °C
0 to 500 °C
0.95 °C
500 to 1750 °C
0.48 °C
-20 to 0 °C
1.03 °C
0 to 500 °C
0.93 °C
500 to 1750 °C
0.55 °C
-200 to 0 °C
0.30 °C
0 to 400 °C
0.13 °C
-200 to 0 °C
0.27 °C
0 to 400 °C
0.13 °C