7.9 Manual Calibration Process
Find Quality Products Online at:
Cabling
Scan Mode
AC Adapter
NOTE: 1. The 1523/24 is tested and calibrated using the tests outlined below in the
order indicated. Details on performing each of the tests can be found in the subsequent
section of this document.
2. For test specifications, durations, soak times, and other detailed information, refer
to the specification tables below.
Table 28 1523/24 Accuracy Test Settings and Specifications
Resistance
Reference
Range
(Nominal) (Ω)
L75_OHMS
0
25
75
LO_OHMS
75
100
200
400
MED_OHMS
200
400
10k
40k
GlobalTestSupply
www.
Calibration of Your Reference Thermometer Readout
Soak (Sec)
Samples
20
40
20
40
20
40
20
40
20
40
20
40
20
40
20
40
20
40
20
40
20
40
.com
Manual Calibration Process
Spec
Std Dev
(±Ω)
Spec (±Ω)
0.002
0.0005
0.003
0.00075
0.005
0.00125
0.005
0.00125
0.006
0.0015
0.01
0.0025
0.018
0.0045
0.52
0.13
0.54
0.135
1.5
0.375
4.5
1.125