Trace Oxygen Analyzer

Operation

4.1 Introduction

Once the analyzer has been installed, it can be configured for your
application. To do this you will:
Set system parameters.
Establish a security password, if desired, requiring the
operator to log in.
Establish and start an automatic calibration cycle, if desired.
Calibrate the instrument.
Define the three user selectable analysis ranges, then choose
auto ranging or select a fixed range of analysis, as required.
Set alarm setpoints, and modes of alarm operation (latching,
failsafe, etc).
Before you configure your 3000ZA2G, these default values are in
effect:
Ranges:
Alarm Relays:
Zero:
Span:
Password:
If you choose not to use password protection, the default password
is automatically displayed on the password screen when you start up and
you have access to all functions of the analyzer.
Teledyne Analytical Instruments
LO = 10 ppm, MED = 100 ppm, HI = 1000 ppm.
Auto Ranging: ON
Defeated, 1000 ppm, HI, Not failsafe, Not
latching.
Auto, every 0 days at 0 hours.
Auto, at 000008.00 ppm, every 0 days at 0 hours.
TAI
Operation
27