Endress+Hauser Levelflex FMP51 Operating Instructions Manual

Guided level-radar
Hide thumbs Also See for Levelflex FMP51:
Table of Contents

Advertisement

Operating Instructions
Levelflex FMP51, FMP52, FMP54
Guided Level-Radar
Level and interface measurement in liquids
BA01006F/00/EN/14.12
71206440
Valid as of version
01.00.zz

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Levelflex FMP51

  • Page 1 Operating Instructions Levelflex FMP51, FMP52, FMP54 Guided Level-Radar Level and interface measurement in liquids BA01006F/00/EN/14.12 71206440 Valid as of version 01.00.zz...
  • Page 3: Table Of Contents

    Levelflex FMP51, FMP52, FMP54 Table of contents Table of contents 6.1.6 FMP54: Dimensions of process connection Important document information ..5 and probe ..... . . 32 About this document .
  • Page 4 Table of contents Levelflex FMP51, FMP52, FMP54 12.1.4 Replacement of a device ... . . 93 Integration into a PROFIBUS 12.2 Spare parts ......94 network .
  • Page 5: Important Document Information

    The document types listed are available: • On the CD supplied with the device • In the Download Area of the Endress+Hauser Internet site: www.endress.com ® Download 1.1.3 Safety Instructions (XA) for Levelflex FMP51, FMP52, FMP54 Depending on the approval, the following Safety Instructions (XA) are supplied with the instrument.
  • Page 6: Document Conventions

    Important document information Levelflex FMP51, FMP52, FMP54 51 52 54 Feature Approval Safety Instructions Safety Instructions HART PROFIBUS FOUNDATION Fieldbus CSA C/US DIP Cl.I,II Div.1 Gr.E-G XA00529F XA00570F CSA C/US IS Cl.I,II,III Div.1 Gr.A-G, NI Cl.1 Div.2, Ex ia XA00530F XA00571F CSA C/US XP Cl.I,II,III Div.1 Gr.A-G, NI Cl.1 Div.2, Ex d...
  • Page 7: Electrical Symbols

    Levelflex FMP51, FMP52, FMP54 Important document information Symbol Meaning CAUTION! CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or medium injury. A0011191-EN NOTICE! NOTICE This symbol contains information on procedures and other facts which do not result in personal injury.
  • Page 8: Symbols In Graphics

    Important document information Levelflex FMP51, FMP52, FMP54 Symbol Meaning Indicates additional information. A0011193 Reference to documentation Refers to the corresponding device documentation. A0011194 Reference to page Refers to the corresponding page number. A0011195 Reference to graphic Refers to the corresponding graphic number and page number.
  • Page 9: Basic Safety Instructions

    The manufacturer is not liable for damage caused by improper or non-designated use. Verification for borderline cases: For special measured materials and cleaning agents, Endress+Hauser is glad to provide ► assistance in verifying the corrosion resistance of wetted materials, but does not accept any warranty or liability.
  • Page 10: Operational Safety

    It fulfills general safety requirements and legal requirements. It also conforms to the EC directives listed in the device-specific EC declaration of conformity. Endress+Hauser confirms this fact by applying the CE mark. Endress+Hauser...
  • Page 11: Product Description

    Levelflex FMP51, FMP52, FMP54 Product description Product description Product design 3.1.1 Compact device Levelflex A0012399 Design of the Levelflex å 1 Electronics housing Process connection (here as an example: flange) Rope probe End-of-probe weight Rod probe Coax probe Endress+Hauser...
  • Page 12: Electronics Housing

    Product description Levelflex FMP51, FMP52, FMP54 3.1.2 Electronics housing A0012422 Design of the electronics housing å 2 Electronics compartment cover Display module Main electronics module Cable glands (1 or 2, depending on instrument version) Nameplate I/O electronics module Terminals (pluggable spring terminals)
  • Page 13 Levelflex FMP51, FMP52, FMP54 Product description US Patents EP Patents 6.140.940 6.481.276 6.512.358 1 301 914 6.559.657 1 020 735 6.640.628 6.691.570 6.847.214 7.441.454 7.477.059 1 389 337 7.965.087 Endress+Hauser...
  • Page 14: Incoming Acceptance And Product

    Incoming acceptance and product identification Levelflex FMP51, FMP52, FMP54 Incoming acceptance and product identification Incoming acceptance DELIVERY NOTE 1 = 2 A0013696 A0016870 Is the order code on the delivery note (1) identical to the order code on the product sticker (2)?
  • Page 15: Product Identification

    Are the CD-ROMs (product documentation, operating tool) and documentation present? If required (see nameplate): Are the Safety Instructions (XA) present? If one of the conditions does not comply, contact your Endress+Hauser distributor. Product identification The following options are available for identification of the measuring device: •...
  • Page 16: Nameplate

    Incoming acceptance and product identification Levelflex FMP51, FMP52, FMP54 4.2.1 Nameplate 92 (3.62) Order code: Ser. no.: Ext. ord. cd.: LN = MWP: Lref = Mat.: DeviceID: Dev.Rev.: if modification see sep. label Date: mm (in) 19 18 17 16...
  • Page 17 Levelflex FMP51, FMP52, FMP54 Incoming acceptance and product identification Approval: 51 52 54 ATEX II 1/2G Ex d(ia) IIC T6 ATEX II 1/3G Ex ic(ia) IIC T6 ATEX II 1 D Ex tD IIIC IP6x ATEX II 1/2 D Ex tD IIIC IP6x...
  • Page 18 Incoming acceptance and product identification Levelflex FMP51, FMP52, FMP54 Power Supply, Output 51 52 54 2-wire; 4-20mA HART 2-wire; 4-20mA HART, 4-20mA 2-wire; FOUNDATION Fieldbus, switch output 2-wire. PROFIBUS PA, switch output 4-wire 90-253VAC; 4-20mA HART 4-wire 10,4-48VDC; 4-20mA HART Special version, TSP-no.
  • Page 19 Levelflex FMP51, FMP52, FMP54 Incoming acceptance and product identification Probe: 51 52 54 ..inch, rod 0.63in 316L, 40 inch divisible ..mm, rod 16mm PFA>316L ..inch, rod 0.63in PFA>316L ..mm, rope 4mm 316 ..inch, rope 1/6" 316 ..
  • Page 20 Incoming acceptance and product identification Levelflex FMP51, FMP52, FMP54 Process connection: 51 52 54 8" 150lbs RF, 316/316L flange ANSI B16.5 (CRN) 4" 600lbs RF, 316/316L flange ANSI B16.5 (CRN) 1-1/2" 300lbs RF, 316/316L flange ANSI B16.5 (CRN) AQK 1-1/2" 300lbs, PTFE>316/316L flange ANSI B16.5 AQM 1-1/2"...
  • Page 21 Levelflex FMP51, FMP52, FMP54 Incoming acceptance and product identification Process connection: 51 52 54 Thread ISO228 G1-1/2, 316L (CRN) Thread ISO228 G1-1/2, 200bar, 316L (CRN) Thread ISO228 G1-1/2, 400bar, 316L (CRN) 10K 40 RF, 316L flange JIS B2220 10K 40, PTFE>316L flange JIS B2220 10K 50 RF, 316L flange JIS B2220 10K 50, PTFE>316L flange JIS B2220...
  • Page 22 Incoming acceptance and product identification Levelflex FMP51, FMP52, FMP54 Additional Operation Language: 51 52 54 English German French Spanish Italian Dutch Portuguese Polish Russian Chinese simplified Japanese Korean Czech Application Package: (Multiple options can be selected) 51 52 54 Interface measurement...
  • Page 23 Levelflex FMP51, FMP52, FMP54 Incoming acceptance and product identification Test, Certificate: (Multiple options can be selected) 51 52 54 3.1 Material certificate+PMI test (XRF) internalprocedure, wetted metallic parts, EN10204-3.1 inspection certificate Liquid penetrant test AD2000-HP5-3(PT), wetted/pressure retaining metallic parts, inspection...
  • Page 24 Incoming acceptance and product identification Levelflex FMP51, FMP52, FMP54 Tagging: (Multiple options can be selected) 51 52 54 Tagging (TAG), see additional spec. Bus address, see additional spec. Endress+Hauser...
  • Page 25: Storage, Transport

    Levelflex FMP51, FMP52, FMP54 Storage, Transport Storage, Transport Storage conditions • Permitted storage temperature: –40 to +80 °C (–40 to +176 °F) • Use the original packaging. Transport product to the measuring point WARNING Risk of injury if the hosuing breaks away! Transport the measuring device to the measuring point in its original packaging or at the ►...
  • Page 26: Mounting

    Mounting Levelflex FMP51, FMP52, FMP54 Mounting Mounting dimensions 6.1.1 Dimensions of the electronics housing 90 (3.54) 78 (3.07) A0015132 Housing GT18 (316L); Dimensions in mm (in) å 4 78 (3.07) 90 (3.54) A0015133 Housing GT19 (Plastics PBT); Dimensions in mm (in) å...
  • Page 27: Dimensions Of The Mounting Bracket

    Levelflex FMP51, FMP52, FMP54 Mounting 6.1.2 Dimensions of the mounting bracket 122 (4.8) 140 (5.5) 175 (6.9) 158 (6.2) Ø 42 ... 60 mm (inch) (1-1/4 ... 2) A0014793 Mounting bracket for the electronics housing å 7 Wall mounting Pipe mounting For the "Sensor remote"...
  • Page 28: Fmp51: Dimensions Of Process Connection (G¾,Npt¾) And Probe

    Mounting Levelflex FMP51, FMP52, FMP54 6.1.3 FMP51: Dimensions of process connection (G¾,NPT¾) and probe (4.8) = 100 (4) ø82.5 ø59.35 (ø3.25) (ø2.34) SW36 AF36 ¾ NPT¾ ø10 (0.4) ø4 ø4 ø8 ø22 (ø0.16) (ø0.16) (ø0.31) (ø .87) 2 (0.08) ø29 (ø1.14) ø22...
  • Page 29: Fmp51: Dimensions Of Process Connection (G1½,Npt1½,Flange) And Probe

    Levelflex FMP51, FMP52, FMP54 Mounting 6.1.4 FMP51: Dimensions of process connection (G1½,NPT1½,flange) and probe (4.8) = 100 (4) ø59.35 ø59.35 ø59.35 (ø2.34) (ø2.34) (ø2.34) ø82.5 SW55 SW55 (ø3.25) AF55 AF55 ½ ½ SW10 SW14 AF10 AF14 ø10 (0.4) SW10 AF10 ø16...
  • Page 30 Mounting Levelflex FMP51, FMP52, FMP54 Coax probe; 316L (Feature 060) LN Length of probe Reference point of the measurement Æ95 mm (3.74 in) PEEK 7 mm (0.28 in) 10 mm (0.39 in) 37 mm (1.46 in) 110 mm (4.33 in) Æ45 mm (1.77 in)
  • Page 31: Fmp52: Dimensions Of Process Connection And Probe

    Levelflex FMP51, FMP52, FMP54 Mounting 6.1.5 FMP52: Dimensions of process connection and probe (4.8) = 100 (4) ø82.5 (ø3.25) ø59.8 (ø2.35) ø68.5 ø59.8 ø63.9 ø90.89 (ø 2.7) (ø2.35) (ø 2.52) (ø 3.58) ø16 ø16 ø16 ø16 ø16 (ø 0.63) (ø...
  • Page 32: Fmp54: Dimensions Of Process Connection And Probe

    Mounting Levelflex FMP51, FMP52, FMP54 6.1.6 FMP54: Dimensions of process connection and probe (4.8) SW60 SW60 AF60 AF60 ø82.5 (ø3.25) ø90 ø90 ø90 (ø3.54) ø90 (ø3.54) (ø3.54) (ø3.54) SW14 SW14 SW14 SW14 AF14 AF14 AF14 AF14 M40x1 M40x1 M40x1 M40x1 ø16...
  • Page 33: Mounting Requirements

    Levelflex FMP51, FMP52, FMP54 Mounting Coax probe (Feature 060) LN Length of probe Reference point of the measurement Æ95 mm (3.74 in) PEEK 7 mm (0.28 in) 10 mm (3.94 in) 37 mm (1.46 in) 110 mm (4.33 in) Æ45 mm (1.77 in) 316L 4 mm (0.16 in)
  • Page 34: Applications With Restricted Mounting

    Mounting Levelflex FMP51, FMP52, FMP54 Additional conditions • When mounting in the open, a weather protection cover (1) may be installed to protect the device against extreme weather conditions. • In metallic vessels: Preferably do not mount the probe in the center of the vessel (2), as this would lead to increased interference echoes.
  • Page 35: Notes On The Mechanical Load Of The Probe

    Levelflex FMP51, FMP52, FMP54 Mounting • Levelflex version (see product structure): Feature 600 "Probe Design", Option MB "Sensor remote, 3m/9ft cable, detachable+mounting bracket" (® ä 23) • A connecting cable is supplied with this device version – Length: 3 m (9 ft) –...
  • Page 36 Mounting Levelflex FMP51, FMP52, FMP54 Bending strength of rod probes Sensor Feature 060 Probe Bending strength [Nm] FMP51 AA, AB Rod 8mm (1/3") 316L AC, AD Rod 12mm (1/2") 316L AL, AM Rod 12mm (1/2") AlloyC BA, BB, BC, BD Rod 16mm (0.63") 316L divisible...
  • Page 37: Notes On The Process Connection

    Levelflex FMP51, FMP52, FMP54 Mounting Bending strength of coax probes Sensor Feature 060 Process connection Probe Bending strength [Nm] FMP51 UA, UB Thread G¾ oder NPT¾ Coax 316L, Ø 21,3 • Thread G1½ or NPT1½ Coax 316L, Ø 42,4 • Flange...
  • Page 38 Mounting Levelflex FMP51, FMP52, FMP54 Nozzle mounting with flange mm (in) ø 150 (6) £ A0015122 For FMP52: Use spring washers in order to compensate a possible creep deformation of the PTFE cladding between the tank and the device flange; see figure below.
  • Page 39 Levelflex FMP51, FMP52, FMP54 Mounting Probe Max. nozzle height (= length of the center rod) Option to be selected in feature 060 ("Probe") 300 mm 12 inch FMP52 150 mm 6 inch 300 mm 12 inch Rod extension/centering HMP40 for FMP54 For FMP54 with rope probes the rod extension/centering HMP 40 is available as an accessory (®...
  • Page 40: Securing The Probe

    Mounting Levelflex FMP51, FMP52, FMP54 6.2.5 Securing the probe Securing rope probes A0012609 Sag of the rope: ³ 1 cm per 1m of the probe length (0.12 inch per 1 ft of the probe length) Reliably grounded end of probe...
  • Page 41 Levelflex FMP51, FMP52, FMP54 Mounting ø a ø b ø<25 (1.0) 3 (0.12) » mm (in) A0012607 Probe rod, uncoated Sleeve bored tight to ensure electrical contact between the rod and sleeve! Short metal pipe, e.g. welded in place Probe rod, coated Plastic sleeve, e.g.
  • Page 42: Special Mounting Conditions

    Mounting Levelflex FMP51, FMP52, FMP54 A0012608 Coax probes can be supported at any point of the outer tube. 6.2.6 Special mounting conditions Installation in horizontal and upright cylindrical tanks A0014141 • Any distance from wall, as long as occasional contact is prevented.
  • Page 43 Center washer 3.1 Metallic center washer (316L) for level measurement 3.2 Non-metallic center washer (PEEK, PFA) for interface measurement For information on bypass solutions from Endress+Hauser please contact your Endress +Hauser sales representative. Feature 610 - Accessory mounted Center washer...
  • Page 44 Mounting Levelflex FMP51, FMP52, FMP54 For bypasses with condensate formation (water) and a medium with low dielectric constant (e.g. hydrocarbons): In the course of time the bypass is filled with condensate up to the lower disposal and for low levels the the level echo is superimposed by the condensate echo. Thus in this range the condensate level is measured instead of the correct level.
  • Page 45 Levelflex FMP51, FMP52, FMP54 Mounting • For mechanical reasons, the probe should be installed as vertically as possible. • With inclined installations the probe length has to be adjusted in dependence to the installation angle. – Up to LN = 1 m (3.3 ft): a = 30°...
  • Page 46 Mounting Levelflex FMP51, FMP52, FMP54 Requirements • The dielectric constant of the medium must be at least DC > 7. • The tank wall must be non-conductvie. • Maximum wall thickness (a): – Plastic: < 15 mm (0.6") – Glass: < 10 mm (0.4") •...
  • Page 47 Levelflex FMP51, FMP52, FMP54 Mounting Step Parameter Action Expert ® Sensor ® Sensor properties ® Probe length Select Manual input option. correction ® Confirm probe length Expert ® Sensor ® Sensor properties ® Probe length Enter measured probe length. correction ® Present probe length...
  • Page 48 FMP51 and FMP54 are a perfect replacement for a conventional displacer system in an existing displacer chamber. Endress+Hauser offers flanges that suit Fischer and Masoneilan displacer chamber for this purpose (special product for FMP51; feature 100, options LNJ, LPJ, LQJ for FMP54).
  • Page 49 Levelflex FMP51, FMP52, FMP54 Mounting A0014153 Flange of the displacer chamber Planning instructions: • In normal cases, use a rod probe. When installing into a metallic displacer chamber up to 150 mm, you have all the advantages of a coax probe.
  • Page 50: Mounting The Device

    Mounting Levelflex FMP51, FMP52, FMP54 Mounting the device 6.3.1 Required mounting tools • For mounting thread 3/4": Hexagonal wrench 36 mm • For mounting thread 1-1/2": Hexagonal wrench 55 mm • To shorten rod or coax probes: Saw • To shorten rope probes: –...
  • Page 51: Fmp54 With Gas Phase Compensation: Mounting The Probe Rod

    Levelflex FMP51, FMP52, FMP54 Mounting 4 / 6 mm 0.16 / 0.25 inch 5/15 Nm A0012453 1. Loosen the 3 Allen set screws using an Allen key AF3 (for 4mm ropes) or AF4 (for 6 mm ropes). Note: The set screws have got a clamping coating in order to prevent accidental loosening.
  • Page 52: Mounting The Device

    Mounting Levelflex FMP51, FMP52, FMP54 Rod probes For rod probes with reference reflection the probe rod is delivered separately and has to be mounted as follows: A0014545 1. Screw the counter nut onto the connection thread (M10x1) of the gland. Take care that the chamfer is oriented to the gland.
  • Page 53: Mounting The "Sensor Remote" Version

    Levelflex FMP51, FMP52, FMP54 Mounting • Tighten with the hexagonal nut only: – Thread 3/4": Hexagonal wrench 36 mm – Thread 1-1/2": Hexagonal wrench 55 mm • Maximum permissible torque: – Thread 3/4": 45 Nm – Thread 1-1/2": 450 Nm •...
  • Page 54 Mounting Levelflex FMP51, FMP52, FMP54 CAUTION The plugs of the connection cable may be damaged by mechanical stress. Mount the probe and the electronics housing tightly before connecting the cable. ► Lay the cable such that it is not exposed to mechanical stress. Minimum bending radius: 100 ►...
  • Page 55: Turning The Transmitter Housing

    Levelflex FMP51, FMP52, FMP54 Mounting 6 Nm 6 Nm = 100 mm (4”) = 100 mm (4”) 6 Nm 6 Nm A0014794 Connecting the cable. There are the following possibilities: å 14 Angled plug at the probe Angled plug at the electronics housing 6.3.6...
  • Page 56: Turning The Display Module

    Mounting Levelflex FMP51, FMP52, FMP54 6.3.7 Turning the display module 3 mm A0013905 1. If present (i.e. for devices with Dust-Ex/DIP approval): Loosen the securing clamp of the electronics compartment cover using an Allen key. 2. Unscrew cover of the electronics compartment from the transmitter housing.
  • Page 57: Electrical Connection

    Levelflex FMP51, FMP52, FMP54 Electrical connection Electrical connection Connection options 7.1.1 PROFIBUS PA / FOUNDATION Fieldbus A0011341 Terminal assignment PROFIBUS PA / FOUNDATION Fieldbus å 15 Without integrated overvoltage protection With integrated overvoltage protection Cable screen: Observe cable specifications (® ä 58)
  • Page 58: Connection Examples For The Switch Output

    Electrical connection Levelflex FMP51, FMP52, FMP54 7.1.2 Connection examples for the switch output For HART devices, the switch output is available as an option. See product structure, feature 20: "Power Supply, Output", option B: "2-wire; 4-20mA HART, switch output" Devices with PROFIBUS PA and FOUNDATION Fieldbus always have a switch output.
  • Page 59: Overvoltage Protection

    < 1.5 pF Nominal arrest impulse voltage (8/20 ms) 10 kA External overvoltage protection HAW562 or HAW569 from Endress+Hauser are suited as external overvoltage protection. For detailed information please refer to the following documents: • HAW562: TI01012K • HAW569: TI01013K Connection data 7.3.1...
  • Page 60: Connecting The Measuring Device

    Electrical connection Levelflex FMP51, FMP52, FMP54 Connecting the measuring device WARNING Explosion hazard! Comply with the relevant national standards. ► Observe the specifications in the Safety Instructions (XA). ► Only use the specified cable glands. ► Check whether the supply voltage matches the specifications on the nameplate.
  • Page 61 Levelflex FMP51, FMP52, FMP54 Electrical connection Connect the cable in accordance with the terminal assignment (® ä 57). 8. When using screened cable: Connect the cable screen to the ground terminal. 9. Screw the cover onto the connection compartment. 10. For instruments with safety pin for the lid: Adjust the safety pin so that its edge is over the edge of the display lid.
  • Page 62: Post-Connection Check

    Electrical connection Levelflex FMP51, FMP52, FMP54 Post-connection check Are cables or the device undamaged (visual inspection)? Do the cables comply with the requirements? Do the cables have adequate strain relief? Are all cable glands installed, firmly tightened and correctly sealed? Does the supply voltage match the specifications on the transmitter nameplate? Is the terminal assignment correct (®...
  • Page 63: Operating Options

    å 18 Display module SD02, push buttons; cover must be open for operation Operating options via CDI interface (= Endress+Hauser Common Data Interface) 2.1 Computer with operating tool (FieldCare) 2.2 Commubox FXA291, connected to the CDI interface of the device 8.1.2...
  • Page 64 Operating options Levelflex FMP51, FMP52, FMP54 PROFIBUS DP ENDRESS + HAUSER PROFIBUS PA Prosonic Micropilot Levelflex A0011275 The complete measuring system consists of devices and: å 20 Segment coupler Computer with Profiboard/Proficard and operating tool (FieldCare) PLC (programmable logic controller) More functions (valves etc.)
  • Page 65: The Operating Menu

    Levelflex FMP51, FMP52, FMP54 Operating options The operating menu 8.2.1 Structure Language Display/operation Parameter 1 Parameter 2 Parameter N Setup Setup parameter 1 Setup parameter 2 Setup parameter N Enter access code Advanced setup Parameter 1 Parameter N Submenu 1...
  • Page 66: Submenus And User Roles

    Operating options Levelflex FMP51, FMP52, FMP54 8.2.2 Submenus and user roles The submenus are designed for different user roles. A user role is defined by typical tasks within the lifecycle of the device. User role Typical tasks Submenu Content/Meaning Operator Tasks in the ongoing process: "Language"...
  • Page 67 Levelflex FMP51, FMP52, FMP54 Operating options WP SIM DISPL Spare part: FMP52X-AB 02.01.03 Dev_R ex works SW update #1 SW update #2 A0013132 1. Unscrew the lid from the compartment for the display and operating module. 2. Slightly turn the display and operating module to remove it from the compartment.
  • Page 68: Display And Operating Module

    Operating options Levelflex FMP51, FMP52, FMP54 Display and operating module 8.3.1 Display appearance User ABC_ DEFG HIJK LMNO PQRS TUVW A0012635 Appearance of the display and operation module for on-site operation å 22 Measured value display (1 value max. size) 1.1 Header containing tag and error symbol (if an error is active)
  • Page 69 Levelflex FMP51, FMP52, FMP54 Operating options Display symbols for the submenus Symbol Meaning Display/operation Is displayed: • in the main menu next to the selection "Display/operation" A0011975 • in the header, if you are in the "Display/operation" menu Setup Is displayed: •...
  • Page 70 Operating options Levelflex FMP51, FMP52, FMP54 Measured value symbols Symbol Meaning Measured values Level A0011995 Distance A0011996 Current output A0011998 Measured current A0011999 Terminal voltage A0012106 Temperature of the electronics or the sensor A0012104 Measuring channels Measuring channel 1 A0012000...
  • Page 71: Operating Elements

    Levelflex FMP51, FMP52, FMP54 Operating options 8.3.2 Operating elements Meaning Minus key For menu, submenu Moves the selection bar upwards in a picklist. A0013969 For text and numeric editor In the input mask, moves the selection bar to the left (backwards).
  • Page 72: Entering Numbers And Text

    Operating options Levelflex FMP51, FMP52, FMP54 8.3.3 Entering numbers and text Numeric editor Text editor A0013941 A0013999 Editing view Display area of the entered values Input mask Operating elements Input mask The following input symbols are available in the input mask of the numeric and text editor:...
  • Page 73 Levelflex FMP51, FMP52, FMP54 Operating options Confirms selection. A0013985 Switches to the selection of the correction tools. A0013987 Exits the input without applying the changes. A0013986 Clears all entered characters. A0014040 Operating symbols in the numeric editor A0016621 A0013985 A0013986 Confirms selection.
  • Page 74: Envelope Curve On The Display And Operating Module

    Operating options Levelflex FMP51, FMP52, FMP54 8.3.4 Envelope curve on the display and operating module In order to assess the measuring signal, the envelope curve and - if a mapping has been recorded - the mapping curve can be displayed:...
  • Page 75: Integration Into A Profibus

    Levelflex FMP51, FMP52, FMP54 Integration into a PROFIBUS network Integration into a PROFIBUS network Overview of the device database files (GSD) Manufacturer ID 17 (0x11) Ident number 0x1558 Profile version 3.02 GSD file Information and files under: • www.endress.com GSD file version •...
  • Page 76 Integration into a PROFIBUS network Levelflex FMP51, FMP52, FMP54 A0015903 Example of software addressing; switch 8 is in position "ON"; the address is defined in the operating menu å 25 (Setup ® Device address) Endress+Hauser...
  • Page 77: Commissioning Via Operating Menu (On-Site Display, Fieldcare)

    Levelflex FMP51, FMP52, FMP54 Commissioning via operating menu (On-site display, FieldCare) Commissioning via operating menu (On-site display, FieldCare) 10.1 Installation and function check Make sure that all final checks have been completed before you start up your measuring point: • Checklist "Post-installation check" (® ä 56) •...
  • Page 78: Setting The Operating Language

    Commissioning via operating menu (On-site display, FieldCare) Levelflex FMP51, FMP52, FMP54 10.4 Setting the operating language 10.4.1 Setting the operating language via the display module A0013637 10.4.2 Setting the language via operating tool (FieldCare) A0015305-EN 10.5 Checking the reference distance This section is only valid for FMP54 with gas phase compensation (product structure: feature 540 "Application Package", option EF or EG).
  • Page 79 Levelflex FMP51, FMP52, FMP54 Commissioning via operating menu (On-site display, FieldCare) After mounting the rod probe in the stilling well or bypass, check and - if necessary - correct the setting of the reference distance in the unpressurized state. Whilst doing so the level should be at least 200 mm below the reference distance L to achieve maximum accuracy.
  • Page 80: Configuration Of A Level Measurement

    Commissioning via operating menu (On-site display, FieldCare) Levelflex FMP51, FMP52, FMP54 10.6 Configuration of a level measurement 20 mA 100% 4 mA A0011360 Configuration parameters for level measurements in liquids å 28 LN = Length of probe R = Reference point of the measurement...
  • Page 81: Configuration Of An Interface Measurement

    If required, lower DCs can be entered into "Setup ® Advanced Setup ® Level ® Medium property". However, for DC<1.6 the measuring range may be reduced. For details please contact Endress+Hauser. For FMP54 with gas phase compensation (product structure: feature 540 "Application Package", option EF or EG) a map must NOT be recorded.
  • Page 82: Configuration Of The On-Site Display

    Commissioning via operating menu (On-site display, FieldCare) Levelflex FMP51, FMP52, FMP54 Schritt Parameter Aktion Beschreibung Setup ® Tank level Select tank level: (® ä 117) • Partially filled (typical selection for measurements in tanks) • Flooded (typical selection for measurements in bypasses) Setup ®...
  • Page 83: Adjustment Of The On-Site Display

    Levelflex FMP51, FMP52, FMP54 Commissioning via operating menu (On-site display, FieldCare) 10.8.3 Adjustment of the on-site display The on-site display can be adjusted in the following menu: Setup ® Advanced setup ® Display (® ä 161) 10.9 Configuration management After commissioning, you can save the current device configuration, copy it to another measuring point or restore the previous device configuration.
  • Page 84: Trouble Shooting

    Trouble shooting Levelflex FMP51, FMP52, FMP54 Trouble shooting 11.1 Trouble-shooting instructions Generel errors Error Possible cause Remedial action Device does not respond. Supply voltage does not match the value Connect the correct voltage. indicated on the nameplate. The polarity of the supply voltage is Correct the polarity.
  • Page 85: Diagnostic Information On Local Display

    Levelflex FMP51, FMP52, FMP54 Trouble shooting Error Possible cause Remedial action Interference echo Carry out mapping over entire probe when the tank is empty (® ä 124). Wrong slope of the level in the Wrong tank type selected. Set the Tank type parameter correctly entire measuring range (®...
  • Page 86 Trouble shooting Levelflex FMP51, FMP52, FMP54 Status signals "Failure" A device error is present. The measured value is no longer valid. A0013956 "Function check" The device is in service mode (e.g. during a simulation). A0013959 "Out of specification" The device is operated: •...
  • Page 87: Calling Up Remedial Measures

    Levelflex FMP51, FMP52, FMP54 Trouble shooting 11.2.2 Calling up remedial measures Message for remedial measures X X X X X X X X X X X X X X 20.50 S441 Curr. output 1 Curr. output 1 (ID:153) S441 2d11h05m55s 1.
  • Page 88: Diagnostic List

    Trouble shooting Levelflex FMP51, FMP52, FMP54 A0014014-EN Calling up remedial measures 1. Navigate to the "Diagnostics" menu. Ã In the "Actual diagnostics" parameter, the diagnostic event is shown with event text. 2. On the right in the display range, hover the cursor over the "Actual diagnostics" parameter.
  • Page 89: Overview Of Diagnostic Events

    Levelflex FMP51, FMP52, FMP54 Trouble shooting About the structure of the remedial measure message (® ä 87) 11.5 Overview of diagnostic events 11.5.1 Sensor element failures Diagnostic event Maintenance instructions Error behavior Code Description F003 Broken probe detected 1. Check map.
  • Page 90: Configuration Failures

    Trouble shooting Levelflex FMP51, FMP52, FMP54 11.5.3 Configuration failures Diagnostic event Maintenance instructions Error behavior Code Description F410 Data transfer 1. Check connection. Alarm 2. Check configuration (languages, outputs). 3. Retry data transfer. F411 Up-/download Up-/download active, please wait. Alarm...
  • Page 91: Event Logbook

    Levelflex FMP51, FMP52, FMP54 Trouble shooting Diagnostic event Maintenance Error behavior instructions Code Description S968 Level limited 1. Check level. Warning 2. Check limit parameters. F970 Linearization 1. Check level. Alarm 2. Check linearization settings. The behavior of this error can be defined in the Setup menu (Setup ® Advanced Setup ® Safety settings ®...
  • Page 92: Filtering The Event Logbook

    Trouble shooting Levelflex FMP51, FMP52, FMP54 11.6.2 Filtering the event logbook Using the Filter options parameter, you can define which category of event messages is displayed in the Events list submenu. Navigation path Diagnostics ® Event logbook ® Filter options Filter categories •...
  • Page 93: Repairs

    12.1.1 Repair concept The Endress+Hauser repair concept assumes that the devices have a modular design and that repairs can be done by the Endress+Hauser service or specially trained customers. Spare parts are contained in suitable kits. They contain the related replacement instructions.
  • Page 94: Spare Parts

    Repairs Levelflex FMP51, FMP52, FMP54 12.2 Spare parts • A few interchangeable measuring device components are identified by a spare part nameplate. This contains information about the spare part. • The connection compartment cover of the device contains a spare part nameplate that includes the following information: –...
  • Page 95: Maintenance

    Levelflex FMP51, FMP52, FMP54 Maintenance Maintenance The measuring device requires no special maintenance. 13.1 Exterior cleaning When exterior-cleaning the device, always use cleaning agents that do not attack the surface of the hosuing and the seals. Endress+Hauser...
  • Page 96: Accessories

    Accessories Levelflex FMP51, FMP52, FMP54 Accessories 14.1 Device-specific accessories Accessory Description Weather protection cover mm (in) A0015466 298.5 (11.8) 273.8 (10.8) 164 (6.46) 255.1 (10) mm (in) A0015472 37.8 mm (1.49 in) 54 mm (2.13 in) The weather protection cover can be ordered together with the device (product structure, feature 620 "Accessory Enclosed", option PB "Weather Protection...
  • Page 97 Levelflex FMP51, FMP52, FMP54 Accessories Accessory Description Mounting bracket for the electronics housing 122 (4.8) 140 (5.5) 175 (6.9) 158 (6.2) Ø 42 ... 60 mm (inch) (1-1/4 ... 2) A0014793 Wall mounting Pipe mounting For the "Sensor remote" device version (see feature 060 of the product structure), the mounting bracket is part of the delivery.
  • Page 98 Accessories Levelflex FMP51, FMP52, FMP54 Accessory Description 115mm; 150-250mm / 6-10" 215mm; 250-350mm / 10-14" 315mm; 350-450mm / 14-18" 415mm; 450-550mm / 18-22" Special version, TSP-no. to be spec. Center washer: Not selected DN40 / 1-1/2", inside-d. = 40-45mm, PPS DN50 / 2", inside-d.
  • Page 99 Levelflex FMP51, FMP52, FMP54 Accessories Accessory Description Mounting kit, isolated A0013586 Insulating sleeve Eye-bolt For reliably insulated fixing of the probe. Maximum process temperature: 150 °C (300 °F) For rope probes 4 mm (1/6 in) or 6 mm (1/4 in) with PA>steel: •...
  • Page 100 Accessories Levelflex FMP51, FMP52, FMP54 Accessory Description Centering disk PEEK Æ 48 to 95 mm 10 (0.39) (1.89 to 3.74 inch) can be used for • FMP51 • FMP54 Ø 48 (1.89) Ø 75 (2.95) Ø 95 (3.74) mm (in) A0014576 The centering disk is suitable for probes with a rod diameter of Ø...
  • Page 101 Levelflex FMP51, FMP52, FMP54 Accessories Accessory Description Centering disk PFA Æ 37 mm (1.46") can be used for • FMP51 • FMP52 • FMP54 mm (inch) A0014577 For 8mm (0.3 in) probes For 12 mm (0.47 in) and 16 mm (0.63 in) probes The centering disk is suitable for probes with a rod diameter of 8 mm (0.31 in), 12 mm (0.47 in) and 16 mm (0.63 in) (also coated rod...
  • Page 102: Communication-Specific Accessories

    For details refer to the document SD01007F. 14.2 Communication-specific accessories Accessory Description Commubox FXA291 Connects Endress+Hauser field devices with CDI interface (= Endress+Hauser Common Data Interface) to the USB interface of a computer. For details refer to Technical Information TI405C/07 14.3 Service-specific accessories Accessory...
  • Page 103: System Components

    Levelflex FMP51, FMP52, FMP54 Accessories 14.4 System components Accessory Description Graphic Data Manager The graphic data manager Memograph M provides information on all the relevant process Memograph M variables. Measured values are recorded correctly, limit values are monitored and measuring points analyzed. The data are stored in the 256 MB internal memory and also on an SD card or USB stick.
  • Page 104: Return

    +Hauser, as an ISO-certified company, is required to follow certain procedures when handling returned products that are in contact with medium. To ensure swift, safe and professional device returns, please read the return procedures and conditions on the Endress+Hauser website at www.services.endress.com/return-material Endress+Hauser...
  • Page 105: Disposal

    Levelflex FMP51, FMP52, FMP54 Disposal Disposal In case of disposal please separate the different components according to their material consistence. Endress+Hauser...
  • Page 106: Overview Of The Operating Menu

    Overview of the operating menu Levelflex FMP51, FMP52, FMP54 Overview of the operating menu Language (® ä 112) Display/operation ® Access status display (® ä 113) Locking status (® ä 113) Format display (® ä 113) Contrast display (® ä 115) Display interval (®...
  • Page 107 Levelflex FMP51, FMP52, FMP54 Overview of the operating menu Advanced conditions (® ä 129) Level unit (® ä 130) Blocking distance (® ä 130) Level correction (® ä 131) Setup ® Advanced setup ® Interface ® Process property (® ä 128) DC lower medium (®...
  • Page 108 Overview of the operating menu Levelflex FMP51, FMP52, FMP54 Setup ® Advanced setup ® Switch output ® Switch output function (® ä 146) Assign status (® ä 146) Assign limit (® ä 147) Assign diagnostic level (® ä 147) Switch-on value (®...
  • Page 109 Levelflex FMP51, FMP52, FMP54 Overview of the operating menu Fail safe time (® ä 159) Fail safe type (® ä 159) Fail safe value (® ä 159) Out value (® ä 160) Out status (® ä 160) Out status HEX (®...
  • Page 110 Overview of the operating menu Levelflex FMP51, FMP52, FMP54 Diagnostics ® Event logbook ® Filter options (® ä 173) Event list (® ä 173) Diagnostics ® Device information ® Device (® ä 175) Serial number (® ä 175) Firmware version (®...
  • Page 111 Levelflex FMP51, FMP52, FMP54 Overview of the operating menu Diagnostics ® Data logging ® Assign channel 1 (® ä 180) Assign channel 2 (® ä 180) Assign channel 3 (® ä 180) Assign channel 4 (® ä 180) Logging interval (®...
  • Page 112: Description Of Device Parameters (Operating Menu)

    Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) • : Marks the navigation path to the parameter via the display and operating module. • : Marks the navigation path to the parameter via an operating tool (e.g. FieldCare).
  • Page 113: Display/Operation" Menu

    Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) 18.1 "Display/operation" menu This menu only appears if the device has a local display. Access status display Display/operation ® Access stat. disp. Navigation Setup ® Advanced setup ® Access stat. disp.
  • Page 114 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Navigation Display/operation ® Format display Setup ® Advanced Setup ® Display ® Format display Description Use this function to select how the measured value is shown on the local display. The display format (size, bar graph etc.) and number of measured values displayed simultaneously (1 to 4) can...
  • Page 115 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Additional information 1 value, max. size A0011948-EN 1 bargraph + 1 value A0012011-EN 2 values A0012013-EN 1 value large + 2 values A0012016-EN 4 values A0012019-EN Contrast display Display/operation ® Contrast display...
  • Page 116: Setup" Menu

    Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Factory setting 30 % Display interval Navigation Display/operation ® Display interval Setup ® Advanced Setup ® Display ® Display interval Description Use this function to set the length of time the measured values are displayed if the values alternate on the display.
  • Page 117 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Selection • m • ft • in • mm Factory setting Tank type Setup ® Tank type Navigation Setup ® Advanced setup ® Interface ® Tank type Description Defines the tank type Selection •...
  • Page 118 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Description Selection of the tank level Selection • Partially filled • Flooded Factory setting Partially filled Additional information • Partially filled: The device searches for 2 echo signals, one for the interface and one for the total level.
  • Page 119 0 to 20 Factory setting Additional information Dielectric constants of important media commonly used in the industry are summarized in the document SD106F, which can be downloaded from the Endress+Hauser web page (www.endress.com). Medium group Setup ® Medium group Navigation Condition Only visible for "Operating mode"...
  • Page 120 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Description Defines the empty calibration E. E is the distance between the reference point (lower edge of the flange or threaded connection) and the minimum level (0%). A0013178 Input range Depending on the selected distance unit and the probe.
  • Page 121 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Description Displays the measured level L (before linearization) A0013194 Additional information The value is displayed in the selected "Level unit" (® ä 130). Interface Setup ® Interface Navigation Description Displays the interface level L...
  • Page 122 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Description Displays the measured distance D from the reference point (lower edge of the flange or threaded connection) to the level. A0013198 Additional Information The value is displayed in the selected "Level unit"(® ä 130).
  • Page 123 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Display options • Strong The evaluated echo exceeds the threshold by at least 10 mV • Medium The evaluated echo exceeds the threshold by at least 5 mV. • Weak The evaluated echo exceeds the threshold by less than 5 mV.
  • Page 124: Mapping" Sequence

    Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 18.2.1 "Mapping" sequence Confirm distance Setup ® Mapping ® Confirm distance Navigation Description Confirmation whether the measured distance matches the actual distance. Depending on the selection, the device automatically determines the range over which the mapping will be recorded.
  • Page 125: Advanced Setup" Submenu

    Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Condition Only visible if the Manual map option has been selected in the Confirm distance parameter. Description Definition of the distance up to which the mapping curve will be recorded.
  • Page 126 If no key is pressed for 10 minutes, or the user goes from the navigation and editing mode back to the measured value display mode, the device automatically locks the write-protected parameters after another 60 s . Please contact your Endress+Hauser Sales Center if you lose your access code Input range 0 to 9 999 Define access code Setup ®...
  • Page 127 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Description Use this function to enter a unique name for the measuring point so it can be identified quickly within the plant. The name is displayed in the header: XXXXXXXXX...
  • Page 128 Unknown Additonal information Dielectric constants of important media commonly used in the industry are summarized in the document SD106F, which can be downloaded from the Endress+Hauser web page (www.endress.com). Process property Setup ® Advanced Setup ® Level ® Process property...
  • Page 129 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Options For "Medium type" = "Liquid" • Fast > 1 m(40")/min • Standard < 1 m(40")/min • Medium < 10 cm(4")/min • Slow < 1 cm(0.4")/min • No filter For "Medium type" = "Solid"...
  • Page 130 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Additional information "Advanced conditions" should only be applied in the "Level" operating mode. For two-phase media the "Emulsion layer" option ensures that always the total level is detected (Example: Oil/condensate application).
  • Page 131 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Additional information No echos are evaluated within the blocking distance UB. Therefore, UB can be used to suppress interference echos within the upper end of the probe. A0013219 Level correction Setup ®...
  • Page 132 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 The "Interface" submenu This submenu is only visible for "Operating mode" = "Interface". Process property Setup ® Advanced Setup ® Interface ® Process property Navigation Description Defines a typical rate of level change Options •...
  • Page 133 80 (Dielelectric constant of water at 20 °C / 68 °F) Additional information Dielectric constants of important media commonly used in the industry are summarized in the document SD106F, which can be downloaded from the Endress+Hauser web page (www.endress.com). Level unit Setup ®...
  • Page 134 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Additional information Echoes from within the blocking distance are not taken into account in the signal evaluation. The upper blocking distance is used • to suppress interference echoes at the top end of the probe (1).
  • Page 135 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) The "Automatic DC calculation" sequence If - in the case of interface measurements - the thickness of the upper medium is known at a certain point of time (e.g. through a manual measurement), the Automatic DC calculation sequence can be used to calculate the dielectric constant of the upper medium.
  • Page 136 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Use calculated DC Setup ® Advanced setup ® Interface ® Autom. DC calc. ® Use calc. DC Navigation Meaning Confirm if the automatically calclualted DC is to be used. Selection •...
  • Page 137 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) The "Linearization" submenu Linearization type Navigation Setup ® Advanced setup ® Linearization ® Linearization type Description Defines the type of linearization Selection • None The level is transmitted to the output without 100% linearization.
  • Page 138 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Description Defines the unit of the linearized value. Selection • Free text • t • lb • ton • kg • impGal • UsGal • cf • cm • dm •...
  • Page 139 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Factory setting Diameter Navigation Setup ® Advanced setup ® Linearization ® Diameter Condition Only visible if one of the following linearization types has been selected: • Horizontal cylinder • Sphere...
  • Page 140 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Navigation Setup ® Advanced setup ® Linearization ® Table mode Condition Only visible if the "Table" linearization type has been selected. Description Defines the method used to enter linearization points into the table.
  • Page 141 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Additional information For Table mode = Manual: Level is a writable parameter. For Table mode = Semi-automatic: Level is a read-only parameter. Customer value Setup ® Advanced setup ® Linearization ® Customer value...
  • Page 142 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 The "Safety settings" submenu Output echo lost Navigation Setup ® Advanced setup ® Safety settings ® Output echo lost Description Defines the output signal in the case of a lost echo.
  • Page 143 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Description Defines the slope of the ramp in the case of a lost echo. 100% A0013269 Delay echo lost Ramp echo lost (positive value) Ramp echo lost (negative value) Input range...
  • Page 144 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 The "WHG confirmation" sequence The "WHG confirmation" sequence is only available for devices with SIL approval (feature 590: "Additional Approval", option LC: "WHG overfill prevention") which are currently not in the WHG-locked state.
  • Page 145 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) The "Probe length correction" sequence The Probe length correction sequence helps to ensure that the end of probe signal within the envelope curve is correctly assigned by the evaluation algorithm. The assignment is correct if the length of probe indicated by the device matches the acutal lenght of the probe.
  • Page 146 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Present length Setup ® Advanced setup ® Probe length corr. ® Pres. length Navigation Description Depending on the parametrization: • In most cases: Displays the measured length of probe (according to the detected end of probe signal).
  • Page 147 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Description Allocates a DI block to the switch output. Options • Off No digital block is allocated. • Digital output 1 • Digital output 2 • Digital output 3 • Digital output 4...
  • Page 148 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Switch-on value Switch-off value Setup ® Advanced setup ® Switch output ® Switch-on value / Switch-off value Navigation Only visible for Switch output function = Limit and Assign limit ¹ Off.
  • Page 149 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Additional information The switching behavior depends on the relative position of the two switch points. Switch-on point > Switch-off point: The output is closed if the measured value exceeds the switch-on point.
  • Page 150 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Prerequisite Only visible for Switch output function = Limit and Assign limit ¹ Off. Description Defines the delay for the switching on of the output. Range of values 0 to 100 s...
  • Page 151 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Invert output signal Setup ® Advanced setup ® Switch output ® Invert output signal Navigation Description Allows to invert the behavior of the switch output. Options • No The behavior of the switch output is as described above.
  • Page 152 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Factory setting Fail safe type Navigation Setup ® Advanced setup ® Analog input 1...6 ® Fail safe type Description Defines the output value of the AI block in case of an error.
  • Page 153 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Display options • Good • Uncertain • Bad Additional information For the meaning of the status refer to BA00034S, "PROFIBUS DP/PA - Guidelines for planning and commissioning" Out status HEX Setup ®...
  • Page 154 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 "Digital input 1" ... "Digital input 4" submenus The Digital input 1 to Digital input 4 submenus are used to configure the DI blocks of the device. Channel Setup ® Advanced setup ® Digital input 1...4 ® Channel Navigation Diagnostics ®...
  • Page 155 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Factory setting Fail safe value Setup ® Advanced setup ® Digital input 1...4 ® Fail safe value Navigation Prerequisite Only visible for Fail safe type = Fail safe value. Description Defines the output value of the DI block in case of an error.
  • Page 156 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Additional information The hexadecimal code give more detailed information on the device state than the Out status parameter. For the meaning of the hexadecimal code refer to BA00034S, "PROFIBUS DP/PA - Guidelines for planning and commissioning".
  • Page 157 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Fail safe type Setup ® Advanced setup ® Analog output 1...4 ® Fail safe type Navigation Description Defines the output behavior of the AO block in the case of an error.
  • Page 158 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Display options • Good • Uncertain • Bad Out status HEX Setup ® Advanced setup ® Analog output 1...4 ® Out status HEX Navigation Diagnostics ® Measured value ® Analog output 1...4 ® Out status HEX Description Displays the hexadecimal status of the output value of the block.
  • Page 159 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Setup ® Advanced setup ® Digital output 1...4 ® Invert Navigation Description Allows to invert the digital output signal (i.e. to swap the states "0" and "1"). Option • Off •...
  • Page 160 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Factory setting Out value Navigation Setup ® Advanced setup ® Digital output 1...4 ® Out value Diagnose ® Measured value ® Digital output 1...4 ® Out value Description Displays the current output value of the block.
  • Page 161 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) The "Display" submenu For operating tools: The Display submenu is only visible if a display module is connected to the device. Format display (® ä 113) Value 1 display Setup ® Advanced setup ® Display ® Value 1 display...
  • Page 162 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Options • x • x.x • x.xx • x.xxx • x.xxxx Factory setting x.xx Value 2 display Setup ® Advanced setup ® Display ® Value 2 display Navigation Description Use this function to select one of the measured values to be shown on the local display. If several measured values are displayed at once, the measured value selected here will be the second value to be displayed.
  • Page 163 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Options • x • x.x • x.xx • x.xxx • x.xxxx Factory setting x.xx Value 3 display Navigation Setup ® Advanced setup ® Display ® Value 3 display Description Use this function to select one of the measured values to be shown on the local display. If more than two measured values are displayed at once, the measured value selected here will be the third value to be displayed.
  • Page 164 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Options • x • x.x • x.xx • x.xxx • x.xxxx Factory setting x.xx Value 4 display Setup ® Advanced setup ® Display ® Value 4 display Navigation Description Use this function to select one of the measured values to be shown on the local display. If four measured values are displayed at once, the measured value selected here will be the fourth value to be displayed.
  • Page 165 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Description Use this function to specify the number of decimal places for measured value 4. This setting does not affect the measuring or computational accuracy of the device. The arrow displayed between the measured value and the unit indicates that the device computes with more digits than are shown on the local display.
  • Page 166 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Additional information Device tag Is defined in the Device tag parameter (® ä 126). Free text Is defined in the Header text parameter (® ä 166). Header text Setup ® Advanced setup ® Display ® Header text...
  • Page 167 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Description Selection of the number format for the representation of measured values. Selection • Decimal • ft-in-1/16" (Only valid for distance units) Factory setting Decimal Decimal places menu Setup ® Advanced setup ® Display ® Dec. places menu...
  • Page 168 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 The "Configuration backup display" submenu The Configuration backup display submenu is only visible if a display module is connected to the device. The configuration of the device can be saved to the display module at a certain point of time (backup).
  • Page 169 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Options • Cancel No action is executed and the user exits the parameter. • Execute backup A backup copy of the current device configuration in the HistoROM (built-in in the device) is saved to the display module of the device.
  • Page 170: The "Diagnostics" Menu

    Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 18.3 The "Diagnostics" menu Actual diagnostics Diagnostics ® Actual diagnos. Navigation Description Use this function to display the current diagnostics message. If two or more messages occur simultaneously, the message with the highest priority is shown on the display.
  • Page 171 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Operating time (® ä 168) Endress+Hauser...
  • Page 172: Diagnsotics List" Submenu

    Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 18.3.1 "Diagnsotics list" submenu Up to 5 diagnostics messages currently pending are displayed in this submenu. If more than 5 messages are pending, the messages with the highest priority are shown on the display.
  • Page 173: The "Event Logbook" Submenu

    Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) 18.3.2 The "Event logbook" submenu Filter options Diagnostics ® Event logbook ® Filter options Navigation Description Use this function to select the category (status signal) whose event messages are displayed in the events list.
  • Page 174 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Additional information User interface Example 1 for display format: I 1091 24d12h13m00s Configuration modified Example 2 for display format: S441 01d4h12min30s Current output 1 HistoROM A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
  • Page 175: Device Information" Submenu

    Use this function to view the serial number of the device. It can also be found on the nameplate. Uses of the serial number • To identify the device quickly, e.g. when contacting Endress+Hauser. • To obtain specific information on the device using the Device Viewer: www.endress.com/deviceviewer...
  • Page 176 Uses of the order code • To order an identical spare device. • To identify the device quickly and easily, e.g. when contacting Endress+Hauser. Display Max. 20-digit character string comprising letters, numbers, punctuation marks or +, - Extended order code 1-3 Diagnostics ®...
  • Page 177: Measured Value" Submenu

    Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) 18.3.4 "Measured value" submenu Distance (® ä 121) Level linearized Diagnostics ® Measured val. ® Level linearized Navigation Description Displays the linearized level. Interface distance (® ä 122) Interface linearized Diagnostics ®...
  • Page 178 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 "Analog input 1" ... "Analog input 6" submenu Channel (® ä 151) Out value (® ä 152) Out status (® ä 152) Out status HEX (® ä 153) "Digital input 1" ... "Digital input 4" submenu Channel (®...
  • Page 179 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Out status (® ä 160) Out status HEX (® ä 160) Endress+Hauser...
  • Page 180: Data Logging" Submenu

    Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 18.3.5 "Data logging" submenu Assign channel 1-4 Diagnostics ® Data logging ® Assign channel 1 Navigation Diagnostics ® Data logging ® Assign channel 2 Diagnostics ® Data logging ® Assign channel 3 Diagnostics ®...
  • Page 181 Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) Description Definition of the logging interval t for data logging. This defines the interval between the individual data points in the data log, and thus the maximum loggable process time T = 1000 ×...
  • Page 182 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Description Use this function to view the measured value trend for the logging channel in the form of a chart. A0013859 • x-axis: depending on the number of channels selected displays 250 to 1000 measured values of a process variable.
  • Page 183: Simulation" Submenu

    Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) 18.3.6 "Simulation" submenu Assignment of measured variable Diagnostics ® Simulation ® Assign. meas. var. Navigation Description Use this function to select a process variable for the simulation process that is activated. The display alternates between the measured value and a diagnostics message of the "function check"...
  • Page 184 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Description Use this parameter to activate or deactivate the simulation of the switch output. Options • Off • On Factory setting Off (i.e. no simulation) Switch status Navigation Diagnostics ® Simulation ® Switch status Prerequisite Only visible for Switch output simulation = On.
  • Page 185: The "Device Check" Submenu

    Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) 18.3.7 The "Device check" submenu Start device check Diagnostics ® Device check ® Start device check Navigation Description Start of a device check. Selection • No No device check is performed.
  • Page 186 Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54 Navigation Diagnostics ® Device check ® Level signal Conditions Only visible if a device check has been performed. Description Displays the result of the device check for the level signal.
  • Page 187: Device Reset" Submenu

    Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu) 18.3.8 "Device reset" submenu Device reset Diagnostics ® Device reset ® Device reset Navigation Description Use this function to reset the device configuration - either entirely or in part - to a defined state.
  • Page 188: Index

    Index Levelflex FMP51, FMP52, FMP54 Index DC lower medium (Parameter) ....132 DC value ....... . . 119 Access status display (Parameter) .
  • Page 189 Levelflex FMP51, FMP52, FMP54 Index Distance upper connection (Parameter) ... . . 118 Interface distance (Parameter) ....122 Interface linearized (Parameter) .
  • Page 190 Index Levelflex FMP51, FMP52, FMP54 Hardware locking ......66 Securing coax probes ......41 Software locking .
  • Page 191 Levelflex FMP51, FMP52, FMP54 Index Use calculated DC (Parameter) ....136 User roles ........66 Value 1 display (Parameter) .
  • Page 193 Erklärung zur Kontamination und Reinigung Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the RA# clearly on the outside of the box. If this procedure is not followed, it may result in the refusal of the package at our facility.
  • Page 194 www.endress.com/worldwide *71206440* BA01006F/00/EN/14.12 71206440 CCS/COSIMA 71206440...

This manual is also suitable for:

Levelflex fmp52Levelflex fmp53

Table of Contents