Invacare TDX SP User Manual page 78

Hide thumbs Also See for TDX SP:
Table of Contents

Advertisement

Invacare® TDX® SP
SYMPTOM
PROBABLE
CTRL NOT
The CMPJ or
CONNECTED
Display module is
not communicating
with the control
module (Error code
E200).
CTRL COM FAULT
The controller has
displays and the
determined a fault
wheelchair drives
during a previous
slowly.
turn-off process
(Error code E41).
MISSING
One or more
CONFIGURATION
devices have
displays at power
been removed or
up.
disconnected from
the wheelchair.
ATTENDANT
The Proportional
ACTIVE and
or Digital Attendant
displays.
control is active and
can be used to drive
the chair.
78
CAUSE
SOLUTIONS
Check the
connections
between the joystick
or display and the
controller. Turn the
power off and then
back on. Replace
the controller if
necessary.
Turn the wheelchair
off and back on.
Reconnect the
device.
This is normal
behavior.
SYMPTOM
PROBABLE
CAUSE
Batteries draw
Battery failure.
excessive current
Electrical
when charging.
malfunction.
Battery indicator
Battery failure.
flashes the charge
Malfunctioning
level is low -
battery charger.
immediately after
Electrical
recharge.
malfunction.
Battery indicator
Batteries not
flashes the charge
charged.
level is low - too
Weak batteries.
soon after being
recharged.
SOLUTIONS
Have batteries
checked for shorted
cell. Replace if
necessary.
Contact
Dealer/Invacare
for service.
Check batteries
for shorted cell.
Replace if necessary.
Contact
Dealer/Invacare
for Service.
Contact
Dealer/Invacare
for Service.
Have charger
checked.
Replace batteries
if necessary. Refer
to 8.12 Replacing
Batteries and/or
Battery Cables, page
67.
1143190-N-03

Advertisement

Table of Contents
loading

Table of Contents