68
WARNING!
Risk of Death or Serious Injury
Failure to observe these warnings can cause an electrical
short resulting in death, serious injury, or damage to
the electrical system.
– The POSITIVE (+) RED battery cable MUST connect
to the POSITIVE (+) battery terminal(s)/post(s).
– The NEGATIVE (-) BLACK battery cable
MUST connect to the NEGATIVE (-) battery
terminal(s)/post(s).
– NEVER allow any of your tools and/or battery cable(s)
to contact BOTH battery post(s) at the same time.
An electrical short may occur and serious injury or
damage may occur.
– Install protective caps on positive and negative battery
terminals.
– Replace cable(s) immediately if cable(s) insulation
becomes damaged.
– DO NOT remove fuse or mounting hardware from
POSITIVE (+) red battery cable mounting screw.
1.
Remove the batteries from the wheelchair. Refer to 8.11
Removing/Installing the Batteries From/Into the Wheelchair,
page 65.
2.
Cut the tie-wraps A that secure the battery terminal covers B
to the battery terminals.
3.
Slide the RED battery terminal cover back on the red battery
cable to expose the POSITIVE battery terminal C.
4.
Slide the BLACK battery terminal cover back on the Black
battery cable to expose NEGATIVE battery terminal D.
Remove the locknut E that secures the bracket F of the
5.
POSITIVE battery cable to the POSITIVE (+) battery post of the
battery.