Fluke 4180 Technical Manual page 41

Precision infrared calibrator
Hide thumbs Also See for 4180:
Table of Contents

Advertisement

Precision Infrared Calibrator
Digital Communication Interface
SOUR:STAB:LIM[?] [n]
Read or set the control temperature stability limit, °C or °F where "n" is a positive real value.
Range = {0.01 to 5.0 (°C)}; Default: 0.1 (°C) (Model 4180) 0.4 (°C) (Model 4181)
Read Example: SOUR:STAB:LIM?
Response: 0.1
Set Example: SOUR:STAB:LIM 0.09
SOUR:STAB:TEST?
Read the temperature stability test results. Stable = 1; Unstable = 0
Example: SOUR:STAB:TEST?
Response: 0
A value of 0 is returned if the controller is not stable at the current set-point. Otherwise a value of 1 is
returned if the controller is stable at the current set-point.
SYST:BEEP:IMM
Beep the system beeper.
Example: SYST:BEEP:IMM
The system beeper should make an audible sound in response to this command.
SYST:BEEP:KEYB[?] [n]
Read or set the keyboard beep function, 0=Off, 1=On. Default: 1
Read Example: SYST:BEEP:KEYB?
Response: 1
Set Example: SYST:BEEP:KEYB 1
Turns the keyboard beep function on or off.
SYST:COD:VERS?
Read the main code version.
Example: SYST:COD:VERS?
Response: 1.10
Provides the user with the version of the main processor code.