Fluke 4180 Technical Manual page 33

Precision infrared calibrator
Hide thumbs Also See for 4180:
Table of Contents

Advertisement

Precision Infrared Calibrator
Digital Communication Interface
• <bool> indicates a Boolean value (0 or 1) is required. The mnemonics OFF and ON are also
accepted for 0 and 1, respectively.
• <conv> indicates a conversion mnemonic is required.
• <param> indicates a parameter name is required.
• <seri> indicates a serial number is required.
• <unit> indicates a temperature unit is required.
• <temp> indicates a temperature °C/F is required.
• <pass> indicates a password is required.
• <port> indicates a port number is required.
• <label> indicates an eight character label is required.
• <year> indicates a four digit number is required.
• <month> indicates a one or two digit number is required.
• <day> indicates a one or two digit number is required.
• <hour> indicates a one or two digit number is required.
• <minute> indicates a one or two digit number is required.
• <second> indicates a one or two digit number is required.
• <baud> indicates a valid baud number is required.
*CLS
Clear the status registers.
Example: *CLS
This command has no response.
Clears all status registers (for example, events, and operations).
*IDN?
Read the product information (Manufacturer, Model Number, Serial Number, and Firmware Version).
Example: *IDN?
Response: FLUKE,4180,A79002,1.00
OUTP:DATA?
Read the main heat output percent.
Example: OUTP1:DATA?
Response: 18.0
This command returns the current main zone heater duty cycle.
29