Fluke 4180 Technical Manual page 40

Precision infrared calibrator
Hide thumbs Also See for 4180:
Table of Contents

Advertisement

4180, 4181
Technical Guide
SOUR:SENS:BLOC?
Reads the target temperature (uncompensated sensor temperature MENU|VIEW TEMP|BLOCK
TEMP) in °C or °F.
Read Example: SOUR:SENS:BLOC?
Response: 24.091
SOUR:SENS:DATA?
Reads the apparent temperature, in °C or °F.
Read Example: SOUR:SENS:DATA?
Response: 24.091 (current apparent temperature)
SOUR:SPO[?] [n]
Set the control set-point, °C or °F, where "n" is a real value with acceptance limits based on the model.
Parameter
4180
4181
Read Example: SOUR:SPO?
Response: 50.000
Set Example: SOUR:SPO 100.00
SOUR:STAB:BEEP[?] [n]
Read or set the stability alert (beep) enable where "n" is a value 0 or 1. [0] is disable, [1] is enable
beep. Default: 1 (Enable Beep)
Read Example: SOUR:STAB:BEEP?
Response: 1
Set Example: SOUR:STAB:BEEP 0
Enable or disable the audible stability alert.
SOUR:STAB:DAT?
Read the control temperature stability, °C or °F.
Example: SOUR:STAB:DAT?
Response: 0.306
The controller stability is returned.
36
Min
Max
-15.00
120.00
25.00
500.00
Default
25.0
25.0