Fluke 4180 Technical Manual page 42

Precision infrared calibrator
Hide thumbs Also See for 4180:
Table of Contents

Advertisement

4180, 4181
Technical Guide
SYST:COMM:SER:BAUD[?] [<baud>]
Read or set serial interface baud rate where baud is a standard baud rate value.
Range baud = {1200, 2400, 4800, 9600, 19200, and 38400}; Default: 9600
Read Example: SYST:COMM:SER:BAUD?
Response: 2400
Set Example: SYST:COMM:SER:BAUD 9600
SYST:COMM:SER:LIN[?] [n]
Set serial interface linefeed enable, where n is a value 1 or 0. [0] = LF OFF, [1] = LF ON; Default: 0
(OFF)
Read Example: SYST:COMM:SER:LIN?
Response: 0
Set Example: SYST:COMM:SER:LIN 1
This command enables or disables line feed.
SYST:DEC:FORM[?] [n]
Read or set the decimal format, where n is period [0], comma [1]. Default: 0 (Period)
Read Example: SYST:DEC:FORM?
Response: 0
Set Example: SYST:DEC:FORM 1
SYST:ERR?
Read the most recent error from the error queue.
Example: SYST:ERR?
Response: command protected
This command response reports the errors in the error queue.
SYST:KLOC[?] [n]
This command is unconditionally protected and requires a password to set it.
Read or set the keypad lockout; [0] = unlock, and [1] = lock. Default: 0 (Unlock)
Read Example: SYST:KLOCK?
Response: 1
Set Example: SYST:KLOC 1
This command locks or unlocks the system keypad providing control only through the serial interface
(RS-232 port) or the keypad.
38
Note