Info menu

Event Summary

Used Hours

Device Information

Reminders

16
Event summary displays the summary of three
types of events: changes in settings, alarms and
system events (eg, connection of ResMed USB
stick). There are up to 200 events of each type,
displayed in chronological order, with the most
recent event displayed at the top by default.
Note: When the capacity of 200 logged events is
reached, the oldest log record is overwritten.
Used Hours during the last seven days of
treatment are displayed in a bar graph and can be
compared to the data of the last 365 days.
This screen shows the serial number (displayed
as Device SN), device number, software version
and other component versions. Data on this
screen may be requested for servicing or as part
of problem-solving by a technician.
The Clinician uses the Reminders menu to alert
you to specific events; for example, when to
replace the mask, when to replace the filter and
so on. The reminder appears in yellow as the date
approaches (within 10% of the reminder period).
The reminder also displays when the device is
powered on. You can clear a reminder message
by selecting 'Reset', which clears the current
reminder date to OFF, or displays the next pre-
set Reminder date.