Download Print this page
ResMed Stellar 150 Quick Quide

ResMed Stellar 150 Quick Quide

Non-invasive ventilation with single leak circuit in st mode
Hide thumbs Also See for Stellar 150:

Advertisement

Quick Links

Quick guide to initiate Non-Invasive Ventilation
with single leak circuit in ST mode with Stellar
Covid19 epidemic period
April 2020

Advertisement

loading

Summary of Contents for ResMed Stellar 150

  • Page 1 Quick guide to initiate Non-Invasive Ventilation with single leak circuit in ST mode with Stellar Covid19 epidemic period April 2020...
  • Page 2 • Please note that printed materials may not be the latest version available for download | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 3 (Place the arrows towards the red plug) Patient circuit conected to the ventilator Leak valve with Antibacterial/Antiviral filter anti-asphyxia protection | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 4 Example of single limb with leak circuit assembly | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 5 More difficult inspiratory triggers for the patient (adjust) Moisture accumulation in the filter (HME), which may require changing the filter several times a day | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 6 To turn off the ventilator • Press the power switch on the back of the device, then follow the instructions on the screen | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 7: About The Control Panel

    About the control panel | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 8: Accessing Clinical Mode

    • Proceed in the same way to lock the device (back to the patient mode) | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 9 • Select the mask type Trach (select Setup menu, then Clinical Settings, then Advanced Settings) | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 10 Perform the Learn Circuit | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 11 • Access the Setup menu • Adjust the setting according to the patients needs Then • Press to start ventilation | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 12: Accessing The Alarms Menu

    Accessing the Alarms Menu | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 13: Adding Supplemental Oxygen

    Start ventilation. Turn on the oxygen source and adjust to the desired flow rate Up to 30 L/min can be added | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 14 Attach a new FiO monitoring sensor (as shown below) Perform the sensor calibration. Note: Turn off the oxygen before calibrating the sensor. | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 15 Possibility to set a Low and High FiO2 alarm during ventilation (default values Low 20%, High 100%, default setting off) | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 16 • FR : 20 Mask : Full Face • NB : Using another type of interface may change these results | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 17 Monitoring menu (8 screens) • Viewing ventilation data : During ventilation the screens can be viewed by pressing and turning the Push Dial | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...
  • Page 18 ST mode with Stellar This document is based on information available at the time of publication Document No: 110297. Copyright ResMed | © 2020 ResMed. 04/20 - COVID-19 - This document is based on information available at the time of publication...