Maintenance Introduction; Test Equipment - Agilent Technologies 4349B Manual

4-channel high resistance meter
Hide thumbs Also See for 4349B:
Table of Contents

Advertisement

Maintenance
Introduction
This chapter provides information on how to maintain the 4349B. The 4349B maintenance
consists of performance tests and functional tests.

Test Equipment

Table 9-1 lists the recommended equipment for performing 4349B maintenance. A DC
voltmeter and a DC power supply are required to obtain the reference voltage. The voltage
calibrators listed in Table 9-2 can be used as a substitute for them.
Note
The reference DC voltage is required to test the 4349B current measurement
accuracy and resistance measurement accuracy. To obtain the reference
voltage, the DC power supply requires the DC voltmeter which monitors the
power supply output voltage, because the voltage accuracy of the power supply
is not good enough for the performance tests. On the other hand the voltage
calibrators require no DC voltmeter because of their much greater accuracy.
Equipment
DC Voltmeter
DC Power Supply
RC Box
SMU Test Adapter
Adapter
Cable
Handler Interface
Tester
P: Performance Tests, F: Functional Tests
1
Table 9-1. Required Equipment
Requirements
Voltage Range: 1 V to 100 V
6
Accuracy: <
0.1%
Voltage Range: 1 V to 100 V
No substitute
No substitute
Triaxial Tee
Dual Banana-BNC(m) Cable
61 cm BNC(m)-BNC(m) Cable
No Substitute
Table 9-2. Substitute Equipment for Reference Voltage
3245A w/ Opt 002
DATRON 4000
DATRON 4000A
DATRON 4200
DATRON 4700
DATRON 4707
DATRON 4708
FRUKE 5700
Recommended
3458A
6634A
16340A
PN 04145-65001
PN 1250-1551
PN 11001-60001
PN 8120-1839
PN 04339-65007
Model Number
Qty.
Model
1
1
1
1
1
1
1
1
Maintenance
9
1
Use
P
P
P
F
P
P
F
F
9-1