Making a Measurement

This section discusses the methods for making measurements after con guring for
measurement.

Triggering a Measurement

The 4349B has four trigger source modes: Internal, Manual, External, or Bus. The Trigger
annunciator(
) shows which trigger source is selected.
9
Note
When the bus trigger mode is selected, none of the Trigger annunciator(
ON. The bus trigger mode can be set by GPIB commands only.
To Trigger Internally
1. Press
until the Int annunciator(
To Trigger Manually
1. Press
until the Man annunciator(
2. Press
to trigger a measurement.
To Trigger Externally
1. Connect an external trigger source to the Ext Trigger terminal on the 4349B's rear panel.
2. Press
until the Ext annunciator(
3. Apply a TTL level trigger signal to trigger a measurement. (Refer to \External Trigger
Terminal" in Chapter 3 for trigger speci cations.)

To Use the Comparator Function

The comparator function can be used to sort DUTs according to user set parameter limits.
To set the limit values:
1. Press
to select the channel to set.
2. Press
.
Enter the lower limit value, and press
2-12
Operating the 4349B
) is ON.
9
) is ON.
9
) is ON.
9
.