Agilent 4339B High Resistance Meter
User's Guide
Agilent Part No. 04339-90041
Printed in JAPAN January 2001
Fifth Edition

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies 4339B

  • Page 1 Agilent 4339B High Resistance Meter User's Guide Agilent Part No. 04339-90041 Printed in JAPAN January 2001 Fifth Edition...
  • Page 2 This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of the Agilent Technologies. Agilent Technologies Japan, Ltd.
  • Page 3 4339B Manual Printing History March 1996 First Edition (part number: 04339-90021) : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :...
  • Page 4 In addition it violates safety standards of design, manufacture, and intended use of the instrument. The Agilent Technologies assumes no liability for the customer's failure to comply with these requirements. 4339B is designed for use in INSTALLATION CATEGORY II according to IEC Note 61010-1 and POLLUTION DEGREE 1 according to IEC 61010-1 and IEC 60664-1.
  • Page 5 4339B Dangerous Procedure Warnings Warnings , such as the example below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. Dangerous voltages, capable of causing death, are present in this Warning instrument. Use extreme caution when handling, testing, and adjusting...
  • Page 6 Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from another country.
  • Page 7 4339B Exclusive Remedies The remedies provided herein are buyer's sole and exclusive remedies. Agilent Technologies shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any other legal theory. Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products.
  • Page 8 4339B Safety Symbols General de nitions of safety symbols used on equipment or in manuals are listed below. Instruction manual symbol: the product is marked with this symbol when it is necessary for the user to refer to the instruction manual.
  • Page 9 4339B 4339B High Resistance Meter at a Glance. touch the UNKNOWN terminals or the electrodes of the accessory, Warning when the High Voltage indicator is ON.
  • Page 10 4339B In User's Guide Chapter 1, Preparation for Use For initial turn on of the 4339B Chapter 2, Operating the 4339B Basic measurement operation Getting acquainted with the 4339B|for beginners Handy reference for common measurement tasks|for all users Chapter 3, Measurement Examples...
  • Page 11: Table Of Contents

    In This Chapter ....... To Reset 4339B to its Default Settings .....
  • Page 12 Printing ........2-14 Disabling Printing ......2-14 To Use Measurement Sequence Function |Controlling charge-measurement in a sequence .
  • Page 13 Figures 2-1. Printer Output ....... 2-14 Tables 1-1. Power Voltage Selector Setting ..... . . Contents-3...
  • Page 15: Preparation For Use

    Preparation for Use In This Chapter Before turning the 4339B ON, you must rst set the 4339B to match the available power LINE voltage. If the 4339B's power LINE voltage and frequency are properly set and ready to use, you can skip this chapter.
  • Page 16: To Set Power Line Frequency

    1. Connect the power cable to the power cord receptacle on the rear panel. 2. Push the LINE switch in. The 4339B will emit a beep and perform the self test. (If any message is displayed, see \Error Messages" back of .) The 4339B...
  • Page 17: Operating The 4339B

    Operating the 4339B In This Chapter Basic measurement operations of the 4339B and references are explained. Operating the 4339B...
  • Page 18: To Reset 4339B To Its Default Settings

    UNKNOWN terminals or the electrodes of the accessory, Warning when the High Voltage indicator is ON, the 4339B outputs dangerous voltage of up to 1000 Vdc. Before handling the 4339B or the accessory, turn OFF the test voltage pressing and con rm that the High Voltage indicator is OFF.
  • Page 19: 16117B Low Noise Test Lead

    Three type modules are available. For detail see \Accessories Available" later in this chapter. * Measuring a high capacitance DUT keeping good S/N ratio, change the short bar to an appropriate resistor. For detail, see page 3-6 of 16339A Component Test Fixture Operation and Service Manual Operating the 4339B...
  • Page 20: To Perform Calibration|Canceling Internal Measurement Errors

    To Perform Calibration|Canceling internal measurement errors 1. Press 2. Select using and press . The 4339B will perform the calibration. ExecCal After the calibration is completed with the message , the 4339B will Calibration Complete return to the measurement display.
  • Page 21: To Perform Open Correction |Canceling The Stray Admittance In Parallel With The Dut

    . A source voltage is applied to the test xture, and the V Output indicator will turn ON. Pressing may cause the 4339B to output dangerous voltage, up to Warning 1000 Vdc. Do touch the UNKNOWN terminals or the electrodes of the accessory when the V Output indicator is ON.
  • Page 22: To Connect Dut

    2. Select the desired parameter using and press :Resistance, :Current, :Volume resistivity :Surface resistivity If the 16008B resistivity cell is connected to the 4339B, to change volume and Note surface resistivity, switch the volume/surface selector on the resistivity cell. Operating the 4339B...
  • Page 23: Setting The Parameters For Resistivity Measurement

    1. Select using and press ResCell 2. Select the electrode size that you want to use ( , or ) using , and 26mm 50mm 76mm press 3. Press 4. Select and press to exit the menu. Exit Operating the 4339B...
  • Page 24: To Select Measurement Range

    . The measurement range setup menu is displayed. 2. Press until the desired range is displayed. Or, input the current value to be measured using the numeric ENTRY keys, and the 4339B will select the optimum measurement range setting. 3. Press .
  • Page 25: To Set Averaging Rate|Stabilizing The Measurement Result

    2. Enter the desired trigger delay time using the numeric ENTRY keys. (For example, to set 0.5 s, press .) You can set the trigger delay time from 0 s to 9.999 s. 3. Press to set the value and to exit. Operating the 4339B...
  • Page 26: To Use Deviation Measurement Function

    4. Press to enter the value. Selecting the Deviation Mode 5. Select using and press ModeSet 6. Select the desired mode using and press mode: Measured value Reference 1ABS % mode: (Measured value Reference)/Reference 100 % 2-10 Operating the 4339B...
  • Page 27: To Use Comparator Function

    The 4339B shows the comparison results using the display, beeper, printer, and 16064B LED Display/Trigger Box. (To use the 16064B, see \Accessories Available" later in this chapter.) For result output to the display, see \To Select Display Mode"...
  • Page 28: To Select Display Mode

    Each display mode shows the result as follows: The Measurement Display mode shows the measurement data: mode Exponent mode Prefix The Comparison Display mode shows the comparison results: The Display OFF mode ( ) does not show any measurement result. DISP OFF 2-12 Operating the 4339B...
  • Page 29: To Select Beeper Mode

    Exit 5. The Cont Chk annunciator( ) turns ON if the contact check function is on. When contact check failed, the 4339B displays (No-Contact). The limit value for the N.C. contact check function is changable. Refer to Operation Manual The OPEN correction function must be performed correctly for a valid contact check.
  • Page 30: To Print Measurement Data

    2. Connect the printer to the 4339B's GPIB port on the rear panel. 3. Turn the printer ON. Printing Set the 4339B to talk only mode (Set the 4339B's GPIB address to 31). 1. Press 2. Press . The Talk Only annunciator( ) turns ON, and the printer begins printing the measurement data.
  • Page 31: To Use Measurement Sequence Function |Controlling Charge-Measurement In A Sequence

    Intvl b. Enter the interval time using the numeric ENTRY keys, and press c. Enter the number of measurement points (equivalent to Memory size), and press 4. Select and press to exit. Exit Operating the 4339B 2-15...
  • Page 32: Starting Measurement Sequence

    4339B's rear panel, and apply a TTL level trigger signal to trigger a measurement. (For details, see the 4339B Operation Manual Note that the 4339B must be set to the external trigger mode to be triggered from an external handler or from the 16064B LED Display/Trigger Box. To Turn OFF Voltage Output Press and con rm the V Output indicator and the High Voltage indicator is turn OFF.
  • Page 33: Reference

    O set-error canceling : OFF Measurement range : Auto OPEN correction data is cleared Accessories Available Test Fixtures and Test Leads Following test xtures and test leads are available for the 4339B for various forms of DUTs. Operating the 4339B 2-17...
  • Page 34 4339B 16008B 16339A 16117B 2-18 Operating the 4339B...
  • Page 35: 16064B Led Display/Trigger Box

    16064B LED Display/Trigger Box The 16064B LED Display/Trigger Box triggers a measurement when its trigger key is pressed, and displays the comparison results using LEDs. It allows you to manually operate the comparator function of the 4339B. Operating the 4339B 2-19...
  • Page 36: Other Topics

    ) turns OFF and the front-panel keys are unlocked. Check that the 16064B LED display/trigger box is connected to the 4339B and it is set to lock out the keys. If so, unlock the keys from the 16064B. is not accepted: Check whether the interlock connector is rmly connected.
  • Page 37: Measurement Examples

    Measurement Examples In This Chapter The 4339B's features are discussed, which you can investigate by trying the typical measurement examples described in this chapter. touch the UNKNOWN terminals or the electrodes of the accessory, Warning when the High Voltage indicator is ON, the 4339B outputs dangerous voltage of up to 1000 Vdc.
  • Page 38: Measuring Insulation Resistance Of Capacitor

    Measurement Parameter : R Measurement Range : Auto range mode Test Voltage : 100 V Use the measurement sequence single mode (measure after charged for 1 minute.) Measurement Procedure 1. Reset the 4339B. a. Press b. Select using and press 2.
  • Page 39 4339B Select using and press . After a while, the calibration will be ExecCal completed with the message \ ". Calibration Complete 4. Set the test voltage. a. Press b. Press to set the test voltage to 100 V. 5. Perform the OPEN correction.
  • Page 40 4339B h. Press i. Select using and press OpenMeas After a while, the OPEN correction will be completed with the message \ Correction ". (If is displayed, see \ To Perform OPEN Correction Complete Out Of Limit |Canceling the stray admittance in parallel with the DUT" in Chapter 2.) j.
  • Page 41: For More Information

    4339B e. Press f. Select using and press Single 10. Press . After charging 1 minute, the measuremen result will be displayed. The following gure shows the typical measurement result display. For More Information To print out the measurement result | See \To Print Measurement Data" in Chapter 2 To select measurement level | See \To Set Test Voltage"...
  • Page 42: Measuring Resistivity Of Insulation Material

    Measurement parameter : Rv( Measurement Range : Auto range mode Test Voltage : 500 V Use the measurement sequence single mode (measure after charged for 1 minute.) Measurement Procedure 1. Reset the 4339B. a. Press b. Select using and press 2.
  • Page 43 4339B 3. Perform calibration. a. Press b. Select using and press ExecCal After a while, the calibration is completed with the message \ ". Calibration Complete 4. Set the test voltage. a. Press b. Press to set the test voltage to 500 V.
  • Page 44 4339B d. Select using and press to select the current measurement mode. e. Press . The Meas Time annunciator( ) will indicates Long . f. Press . The V Output indicator will turn ON. g. Wait until the value has stabilized within 0.5 pA.
  • Page 45: For More Information

    4339B 8. Set the measurement sequence charging time to 1 minute. a. Press b. Select and press Chrg c. Press d. Select and press to exit. Exit e. Press f. Select using and press Single 9. Press . After charging 1 minute, the measurement result will be displayed. The following gure shows the typical measurement result display.