OFFSET
SIZE
UNIT RES
BYTE
BYTE
16
1
17
1
mBar/Torr/P
a
18
1
19
1
20
1
TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01
RANGE
DESCRIPTION
-10 - +10
Setpoint 1 Pressure Value Exponent.
Set the exponent of the exponential value of the R1
setpoint (format X.XE-XX) when it is used for
pressure monitoring (Parameter Buffer, Offset 21,
B
= 4).
0-2
0-99
Setpoint 2 Pressure Value Mantissa.
Set the mantissa of the exponential value of the R2
setpoint (format X.XE-XX) when it is used for
pressure monitoring (Parameter Buffer, Offset 21,
B
= 4). The value set here is divided by 10, that
3-5
means 11 = 1.1
-10 - +10
Setpoint 2 Pressure Value Exponent.
Set the exponent of the exponential value of the R2
setpoint (format X.XE-XX) when it is used for
pressure monitoring (Parameter Buffer, Offset 21,
B
= 4).
3-5
B
= Stop Speed Reading.
0
This function measures the pump speed after a stop
command.
0 = Off
1 = On
B
= Active Stop.
1
This function brakes the pump after a stop
command, by using the motor as a generator.
0 = Off
1 = On
B
- Reserved
2
B
= Start/Stop/Interlock Mode.
3-4
Change the logic of Start/Stop and Interlock inputs.
Ask to Agilent personnel for this option
B
= Vent & Purge Auto/Man
0
0 = Auto (the controller drives purge and vent
valves automatically)
1 = Manual (the user can command the valves by
serial line or profibus)
Technical Information
Profibus Option
WIN
SERIAL
162
176
176
167
107
151
125
171/190
8