Agilent Technologies TwisTorr 305 FS User Manual page 133

Hide thumbs Also See for TwisTorr 305 FS:
Table of Contents

Advertisement

WIN#
READ/
TYPE
WRITE
111
R/W
N
117
R/W
N
120
R/W
N
122
R/W
L
123
124
125
R/W
A/L
126
R/W
N
130
136
R/W
N
137
R/W
N
138
R/W
N
139
R/W
L
140
R/W
N
143
R/W
N
144
R/W
L
147
R/W
N
155
R
N
TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01
MEANING
Analog output type: output voltage
signal proportional to frequency,
power, temperature or pressure
Low Speed frequency value (Hz)
Rotational frequency setting (Hz)
Set vent valve (on/off)
(write if Win125 = 1)
Reserved to Agilent Service
Reserved to Agilent Service
Set the vent valve operation
Vent valve opening delay (1unit = 0.2s)
Reserved to Agilent Service
Gauge Set Point Type
Gauge Set Point Value (Hz/W/s/mbar) 0¸999999
Gauge Set Point delay: time (s)
between the pump
start and the set point check (seconds)
Gauge Set Point Signal Activation Type
the signal can be "high level active" or
"low level active"; "high level active"
indicates the output is ON when the
signal is above the threshold
Gauge Set point hysteresis (in % of
gauge set point value)
External fan configuration
External fan activation
Vent valve opening time (1unit = 0.2s;
0 = infinite)
(Ref. to "vent connector" paragraph)
Actual maximum power limit (W)
(It reads the maximun power
deliverable to the pump)
Technical Information
Examples
ADMITTED VALUES
FACTORY
SET
0 = frequency
1
1 = power
2 = Temperature
3 = Pressure
from 700 to 1010 Hz
700
from low speed
1010
frequency (win 117) to
1010 Hz
On = 1
1
Off = 0
0 = Automatic
0
1 = On command
2 = Auto pump speed
0¸65535
15
(from 0 to 13107s)
0 = Freqency
3
1 = Power
2 = Time
3 = Normal
867
0¸999999
0
0 = high level active
0
1 = low level active
0¸100
2
0 = Always On
0
1 = Controlled by Serial
On = 1
0
Off = 0
0¸65535 (from 0 to
0
13107s)
133/190