Battery backup input
Functional ground
Cable installation
Choose the correct glands for the selected cable type, see Accessories for glands selection.
Ensure shield is mounted under cable clamps - do not make pig tails.
Note
Mains or line-powered PUR cable (no shield) has to be mounted under cable clamps. All cable
glands have to be sufficiently tightened to ensure IP-rating.
SITRANS FM MAG 8000
Operating Instructions, 03/2020, A5E03071515-AH
Male battery connector with black and red wires; black wire is ground. Male battery connector has
to be connected to female connector on backup battery
Black wire with terminal must be connected to the device encapsulation with a screw
Connecting
5.3 Power supply