MAG 8000 CT is approved with OIML R49:2013 and MID for the following installations:
● DN 50 to DN 300 with 0x D
downstream from the sensor.
● DN 350 to DN 600 with 3x D
downstream from the sensor.
For optimal measurement performance, it is recommended to follow the installation guidelines
for the MAG 8000 standard water meter.
Sensor must be completely full of liquid
SITRANS FM MAG 8000
Operating Instructions, 03/2020, A5E03071515-AH
Minimum straight inlet pipe length
Minimum straight outlet pipe length
of straight pipe upstream to the sensor and 0x D
i
of straight pipe upstream to the sensor and 3x D
i
MAG 8000
(recommended)
5x D
i
3x D
i
Installing/Mounting
4.1 Sensor installation
of straight pipe
i
of straight pipe
i
MAG 8000 CT
MAG 8000 CT
DN 50 to DN 300
DN 350 to DN 600
0x D
3x D
i
i
0x D
3x D
i
i