SITRANS FM MAG 8000
Operating Instructions, 03/2020, A5E03071515-AH
Parameter/data type
Totalizer 2 changes date
Verification mode enable
Calibration date
Calibration factor
Gain correction
Sensor offset
Adjustment Factor
Low flow cut-off
Filter time constant
Excitation frequency limit
Excitation frequency sensor
limit
Empty pipe detection ena‐
ble
Empty pipe limit
Factory settings
Fixed parameter or meter data that are not changeable
PS3 production date
and time
No
Calibration date
Sensor-related
Sensor-related
Sensor-related
1
MAG 8000 Standard
(7ME6810) 0.05 % for
FW 3.07 and lower. For
FW 3.09 see table F-1
below.
MAG 8000 CT
(7ME6820) 0.25 % for
FW 3.07 and lower. For
FW 3.09 see table F-2
below.
15 mm/s for FW 3.11 on‐
wards
5 Tau
6.25 Hz for advanced
version and 1/15 Hz for
basic version
Sensor-related
Yes
25000 ohm
Parameter lists
F.4 300-399
Data range
Yes / No
Value is reset to "No" when upload‐
ing device parameters to pc after four
hours
year-month-day T hours:mi‐
nutes:seconds
-2 to 2
0 to 9.9 %
0 to 999 mm/s
1 to 1000
7ME6810 and 7ME6820:
6.25 Hz
(DN 25 to 200 (1" to 8"))
3.125 Hz
(DN 250 to 600 (10" to 24"))
1.5625 Hz
(DN 700 to 1200 (28" to 48"))
7ME6880:
3.125 Hz
(DN 25 to 600 (1" to 24"))
1.5625 Hz
(DN 700 to 1200 (28" to 48"))
Yes / No
0 to 2.15x10
9
147