Using the Calibration Function

© National Instruments Corporation
The
Calibrate_1200
the calibration DACs with the factory constants or the user-defined
constants stored in the EEPROM, or you can perform your own calibration
and directly load these constants into the calibration DACs. To use the
function or the 1200 Calibrate VI for AI calibration,
Calibrate_1200
ground an AI channel at the I/O connector for offset calibration and apply
an accurate voltage reference to another input channel for gain calibration.
You should first configure the ADC for RSE mode, then for the correct
polarity at which you want to perform data acquisition.
To use the
Calibrate_1200
calibration, the DAC0 and DAC1 outputs must be wrapped back and
applied to two other AI channels. You should first configure the AI
circuitry for RSE and for bipolar polarity, then configure the AO circuitry
for the polarity at which you want to perform output waveform generation.
Refer to your software documentation for more details on the
function and the 1200 Calibrate VI.
Calibrate_1200
function and the 1200 Calibrate VI can either load
function or the 1200 Calibrate VI for AO
5-3
Chapter 5
Calibration
NI PCI-1200 User Manual