Enter up to 60 breakpoints via SIPROM T or the remote operation.
• Set the linearization table (60 breakpoints) (Page 162)
Enter the desired number of breakpoints. You can freely position x and y values on the
linearization curve.
Figure 9-1
• Set the spline curve (40 breakpoints) (Page 162)
Enter the minimum and maximum X-value. The x values are distributed evenly on the x axis
according to the desired number of breakpoints. You can freely assign the y values.
Figure 9-2
SITRANS TH100/TH320/TH420/TR320/TR420/TF320/TF420 with 4 to 20 mA/HART
Operating Instructions, 05/2020, A5E41864807-AB
Linearization table
Spline curve
9.4 Assigning parameters using remote operation
Parameter assignment
161