Do Not Service or Adjust Alone ne ne ne
Do Not Service or Adjust Alo
Do Not Service or Adjust Alo
Do Not Service or Adjust Alo
D D D D o Not Subs
o Not Subs
o Not Subst t t t itute Parts or Modify Equipme
o Not Subs

Maintain the System as Recommended

Maintain the System as Recommended
Maintain the System as Recommended
Maintain the System as Recommended
Foll l l l ow All X
Fol
Fol
Fol
ow All X- - - - ray Safety Requirements and Legislation
ow All X
ow All X
Rev. A
Rev. A
Rev. A
Rev. A
Equipment that appears damaged or defective should be
made inoperative and secured against unintended operation
until they can be repaired by qualified service personnel.
See the Lock Out /Tag Out procedure in the Service section
of the on-line help.
Do not attempt internal service or adjustment of this
instrument unless another person, capable of rendering
first aid and resuscitation, is present.
itute Parts or Modify Equipment nt nt nt
itute Parts or Modify Equipme
itute Parts or Modify Equipme
Because of the danger of introducing additional hazards, do
not install substitute parts or perform any unauthorized
modification to the product. Contact your Agilent
Authorized Service Representative for service and repair to
ensure that safety features are maintained.
The recommended preventive maintenance for the system is
documented in the on-line help. Preventive maintenance
should always include testing the emergency shut down
(EMO) switch by pressing it and verifying that all ac and dc
power turns off.
ray Safety Requirements and Legislation
ray Safety Requirements and Legislation
ray Safety Requirements and Legislation
DO NOT operate the equipment unless an X-ray safety
survey has been performed within the last six (6) months,
when the Medalist x6000 is installed at a site, when the
Medalist x6000 has been moved, and immediately following
maintenance that involves the safety interlock subsystem,
the lead shielding or the X-ray source.
Do not operate this equipment without complying to these
recommendations.
Owners/Operators of this equipment are responsible for
compliance with all local legislation regarding X-ray
equipment.
An Operator's Guide and Declaration of Conformity in a
local language is available upon request to Agilent
Technologies, Inc.