Technical specifications
8.2 General technical specifications
8.2.3

Technical specifications of the wall housing

Table 8-6
General
Installation position
Weight
Degree of protection
Table 8-7
Electrical characteristics
Auxiliary power supply
Power consumption
EMC interference immunity (electromagnetic com‐
patibility)
Electrical safety
Power unit cable
Table 8-8
Purging gas pressure
Continuous (recommended)
Continuous (maximum)
Transient (maximum)
Table 8-9
Electrical inputs and outputs
Digital outputs (DO)
Digital inputs (DI)
Ethernet interface (bottom)
Service interface (inside at PU)
82
Wall housing: General technical specifications
Wall housing: Electrical characteristics
Wall housing: Purging gas pressure
Wall housing: Electrical inputs and outputs
Vertical
Approx. 23 kg
IP65 in accordance with EN 60529,
restricted breathing enclosure to EN 50021
100 to 240 V AC (nominal range of use 85 to 264 V)
50 to 60 Hz (nominal range of use 47 to 63 Hz)
280 VA max.
In accordance with the standard requirements of
NAMUR NE21 (05/2006) and EN 61326-1 (2013)
In accordance with EN 61010-1, overvoltage cate‐
gory II
Max. Ø 10 mm
, max. 3 x 2.1 mm
2
30 hPa above ambient
< 100 hPa above ambient pressure
165 hPa above ambient pressure
8, with changeover contacts, parameters can be
freely assigned, e.g. for measuring range identifi‐
cation;
max. load: 24 V AC/DC/1.7 A (total load for all 8
relay outputs in continuous operation max. 160 W),
floating, non-sparking
8, designed for 24 V, isolated, freely parameteriz‐
able, e.g. for autoranging
Ethernet RJ 45, 100 MBit
Ethernet RJ 45, 100 MBit
Devices in explosion-proof design