Siemens SIPROCESS GA700 Compact Operating Instructions page 154

Continuous gas analysis, devices in explosion-proof design
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Available languages

Available languages

Connecting
4.3 Gas connections
Wall-mounted device → bottom
Figure 4-5
Version with pipes
Introduction of toxic, aggressive or flammable gases
Before toxic, aggressive or flammable gases are introduced, carry out a leak test for the pipe
connections.
Note
Pipes
Connect the pipes using suitable threaded joints with the gas connections. The material of the
parts in contact with the gas (threaded joints and gasket material) must be appropriate to the
application.
The sample gas connectors are made of stainless steel no. 1.4404 or Hastelloy no. 2.4819.
The reference gas connector is made of stainless steel no. 1.4404. The gas connectors have
nozzles with an external diameter of 6 mm.
48
Purging gas nozzle 1
OXYMAT 7 analyzer mod‐
ule with:
Purging gas nozzle 2
Example of wall-mounted device, bottom
WARNING
1
Sample gas inlet
2
Sample gas outlet
Compact Operating Instructions, 05/2015, A5E35134047-02
3
not assigned
4
Reference gas inlet
Devices in explosion-proof design

Advertisement

Chapters

Table of Contents
loading

Table of Contents