2-2 WATER BOTTLE ASSEMBLY, MODEL OS-H4

No.
NAME
1
Bottle
2
Water Bottle Cap Assembly
3
Air Pipe Stem
4
Air/Water Hose
5
Air/Water Connector
6
Air/Water- Drain Lever
7
Water Feeding Stem
- If the water bottle assembly has been handled roughly, the water feeding tube inside the Air/Water hose may be
NOTE
disconnected at the A/W connector for the endoscope. To test, remove the cap assembly and using a syringe, inject water into
the water feeding stem. If the water comes out of both the center hole of the A/W connector and the series of holes around the
center hole, the water feeding tube is disconnected. Use another water bottle assembly.
3
2
1
FUNCTION
holds sterile water for procedure. (use up to 2/3's full)
must be firmly secured to bottle to prevent air leakage. Do NOT overtighten the bottle cap.
inserts into video processor water bottle receptacle.
contains (2) independent tubes - 1 for air, 1 for water.
inserts into Air/Water socket of endoscope umbilical connector.
must be set to upright (A/W) position for delivery of air and water.
channel for water to be displaced from bottle and into scope.
6
7
10
4
5