2.4 Instructions and warnings on the housing of Q2

The information on the device must always be observed and followed!
On the device are warnings in
English and / or French.
The type plate does not provide
sufficient space, so there are
additional warning lables on the
housing.
The figure below shows the
possible places.
Figure 2.1:
Locations for warnings

2.5 Safety information about: Application / gases / ambient conditions

The device requires various gases with different risk potential for operation and measurement.
After switching on, an automatic flushing of the internal gas channels is done.
It is not allowed to connect gasses that are flammable or explosive under
exclusion of oxygen(i.e. acetylene (C2H2)).
It is not allowed to connect gas containing more that 3.0 percent of oxygen
(volume-based).
Gases may form an explosive mixture with air, may be toxic, harmful or
dangerous for the environment.
Maximum inlet pressure is not to exceed 0.5MPa(g) (5 bar(g)). The operator
must ensure this!
Only natural and natural gas-like gases (all components gaseous) must be
used.
For other gas types please contact Elster GmbH before usage. For gas
specifications please see chapter 10.