Hybrid And Sensor Cables - Siemens LDS 6 Operating Instructions Manual

In-situ laser continuous gas analyzer
Table of Contents

Advertisement

8.2

Hybrid and Sensor Cables

Table 8- 10
General
Configuration hybrid cable
Cable sheath
Dimensions
- Diameter
- Length
Impact resistance
Maximum tensile strength
Minimum bending radius
Table 8- 11
Climatic conditions
Ambient temperature
Rel. humidity
LDS 6
Operating Instructions, 01/2009, A5E00295894-05
General
Climatic conditions
8.2 Hybrid and Sensor Cables
Two optical fibers and two twisted copper wires in
one cable for 24 V DC. Single-mode optical fiber
configured at both ends with E2000 angle
connectors. Multimode optical fiber configured at
both ends with SMA connectors.
Oil-resistant polyurethane
< 8 mm,
up to 1 000 m
for lengths > 500 m, an external power supply
must be additionally ordered
for installations in Ex zones, non-intrinsically-safe
cables have to be spatially separated from
intrinsically-safe lines, lengths <600 m
200 N/cm
500 N
10 cm
-40 ... +80 °C during operation
< 95 % rel. humidity, above dew point
Technical Data