5.3 Faults Alarm

5.3
Faults Alarm
Any faults in the hardware that makes the analyzer unable to carry out measurements, result
in a "fault alarm". The measured value flashes, and then it is always necessary to take
corrective measures.
The faults listed in the following table lead to a fault alarm message in the logbook. These
can be individually deactivated by using function 88.
Table 5- 3
Fault alarm messages
No.
Error Message
S1
Opto module
S2
Laser Current
S3
Signal Quality
S6
External Fault
S7
Supply Voltage Channel
S8
Supply Voltage Analyzer
S9
Serial EEProm
S11
FPGA Channel
S12
CAN Analyzer
S13
CAN Channel
S14
Data Flow Analyzer
68
Possible Causes
Laser damage.
Leaking reference cell.
Reference cell displacement.
Electronics damage.
Change of laser current due to ageing of laser.
Signal amplitude too high due to high component
concentration.
Signal amplitude too low when measuring
temperature.
External signaling.
Internal power failure.
Internal power failure.
Internal EEProm failure.
Electronics failure in data acquisition unit.
Electronics failure in internal communication.
Electronics failure in internal communication.
Electronics failure in internal communication.
Action
Contact service.
Contact service.
Contact service.
Contact service.
Check external equipment.
Contact service.
Contact service.
Contact service.
Contact service.
Contact service.
Contact service.
Contact service.
Operating Instructions, 01/2009, A5E00295894-05