Hybrid Cable Connection - Siemens LDS 6 Operating Instructions Manual

In-situ laser continuous gas analyzer
Table of Contents

Advertisement

3.3.2

Hybrid Cable Connection

The hybrid cable is connected at the back of the central unit, where its two optical cables and
the supply line for the sensor wire are attached, as shown in the picture below.
Figure 3-1
Keep the fiber end protected, by the protective tube, until it is time for connection. Only
authorized personnel are entitled to remove the protective tube and proceed with the
connecting operation.
1. Hybrid cable holder.
2. E2000 Single Mode fiber connector, angle polished.
3. SMA Multi Mode fiber connector.
4. Sensor voltage supply connection. 24 V/60 mA.
5. Signal output connections.
6. Network connection. Ethernet 10Base-T (RJ-45).
See also
Pin Assignment of LDS 6 (Page 34)
Signal Cable Connection (Page 32)
LDS 6
Operating Instructions, 01/2009, A5E00295894-05
1
LDS 6 cable connections (here only channel 3 is connected)
WARNING
2
Installation Guidelines
3.3 Electrical Connections
3
4
5
6
31