Measuring Ranges; Parameters - Siemens LDS 6 Operating Instructions Manual

In-situ laser continuous gas analyzer
Table of Contents

Advertisement

Operation
4.3 Analyzer Functions
4.3.4

Measuring Ranges

Table 4- 5
41 Define Ranges
41 Define Ranges
Ch1 NH
Start Value End Value
0.000
100.0
mg/Nm
4.3.5

Parameters

Table 4- 6
Parameters
Ch1 NH
50 Response Time
51 Limits
52 Transmission
53 Status Messages
...Continue
50 Response Time
50 Response Time
Ch1 NH
[5.000] Seconds
Actual Measured
Value: 63.28 mg/Nm
3
52
You can use this function to define a measuring range by assigning a start-of-scale value to the
bottom value (4 mA) and a full-scale to the top value (20 mA) of the analog output.
3
3
This display, showing the selection of the parameter functions 50 to 53 appears following the
selection of the parameter functions in the main menu by pressing the fourth soft key
3
("Parameters"). You can branch to the parameter functions 54 to 60 by pressing the fifth soft key
(...Continue).
You can use this function to set various time constants to reduce or suppress noise in the
measured signal.
3
Operating Instructions, 01/2009, A5E00295894-05