Configuration - Siemens LDS 6 Operating Instructions Manual

In-situ laser continuous gas analyzer
Table of Contents

Advertisement

60 Setup Logbook
60 Setup Logbook
Ch1 NH
Clear logbook
Lock logbook
4.3.6

Configuration

All functions of this block are only accessible via the password for level 2 except function 89
-Ethernet On/Off which requires code level 3.
Input Menu
Configuration
Ch1 NH
70 Analog Output
71 Relay Outputs
72 Binary Inputs
73 Communication
...Continue
70 Analog Outputs
70 Analog Output
Ch1 NH
4-20mA (NAMUR)
Suppress negative
measurement values
LDS 6
Operating Instructions, 01/2009, A5E00295894-05
You can use this function to delete or lock logbook entries (see also function 3).
3
Following selection of the configuration functions in the main menu by pressing the fifth soft key
("...Continue"), you can branch to the further configuration functions.
3
You can use this function to switch the mode of the analog output between NAMUR and non
NAMUR . If the NAMUR mode is chosen in this screen, the lower and upper limits of the analog
outputs are 3.8 and 21.5 mA respectively, according to the NAMUR standard. Otherwise the
3
limits are 2 and 21 mA. For both modes function 41 is used to set the limits of the analog output
range to 4 mA (start of scale) and 20 mA (full scale). This means that the wider output range
permits recording of measured values outside the adjusted output range, but the scaling is
identical to the 4 to 20 mA span.
If the function "Suppress negative measurement values" is activated, negative measured values
are set to the analog output value which corresponds to a measuring value of 0. This analog
output value depends on the adjusted measuring range.
If negative measured values have an unfavorable effect on further processing, activate this
function. The correct measured value is still output in the display.
Operation
4.3 Analyzer Functions
55