Technical Specifications - ResMed AirMini Clinical Manual

Hide thumbs Also See for AirMini:
Table of Contents

Advertisement

• The machine should not be used adjacent to or stacked with other equipment. If adjacent
or stacked use is necessary, the machine should be observed to verify normal operation in
the configuration in which it will be used.
• The F20, N20 and P10 connectors for AirMini are fitted with vents. The F20 connector for
AirMini is also fitted with an AAV. The AAV and vents have specific safety functions to
prevent carbon dioxide build-up in the mask and should be kept clean and clear. The
connectors should not be used if the vent or AAV are damaged as they will not be able to
perform their safety functions. The connectors should be replaced if the valves of the vent
or AAV are damaged, distorted or torn.
• The machine is not intended to be operated by persons (including children) with reduced
physical, sensory or mental capabilities without adequate supervision by a person
responsible for the patient's safety.
• The device has not been tested or certified for use in the vicinity of X-ray, CT or MRI
equipment. Do not bring the device within 4 m of X-ray or CT equipment. Never bring the
device into an MR environment.
CAUTION
• Use only ResMed AirMini parts, masks and accessories with the machine. Non-ResMed
AirMini parts may reduce the effectiveness of the treatment, may result in excess carbon
dioxide rebreathing and/or damage the machine. For compatibility information, refer to
www.resmed.com for more information.
• Blocking the air tubing and/or air inlet of the machine while in operation could lead to
overheating of the machine.
• Keep the area around the machine dry, clean and clear of anything (eg, clothes, pillows or
bedding) that could block the air inlet or cover the power supply unit.
• Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturising or
antibacterial soaps or scented oils to clean the machine or air tubing. These solutions may
cause damage and reduce the life of the products.
• Do not insert any USB cable into the machine or attempt to plug the power supply unit into
a USB device. This may cause damage to the machine or USB device.

Technical specifications

Units are expressed in cm H
Power supply unit
AC input range:
DC output:
Typical power consumption:
Peak power consumption:
If power is interrupted during therapy, the machine will automatically restart therapy when power is restored.
AirMini power
Standby power consumption:
Environmental conditions
Operating temperature:
O and hPa. 1 cm H
2
O is equal to 0.98 hPa.
2
100–240V, 50–60Hz, 0.5-0.3A
115V, 400Hz for aircraft use, 0.4A
24V
0.83A
6.3W
27W
1 W
+5°C to +35°C
Note: Under extreme ambient temperature conditions
(40°C) the air and parts of the tubing near the mask may
reach temperatures up to 43°C. These parts of the tubing
are not in contact with the user and the machine remains
safe under these extreme conditions.
English
25