ResMed AirMini Clinical Manual page 18

Hide thumbs Also See for AirMini:
Table of Contents

Advertisement

Parameter
EPR Level
Options
Parameter
Description
SmartStart
If you switch the feature on, the machine will start therapy
automatically when the patient breathes into their mask.
SmartStop
If you switch the feature on, the machine will stop therapy
automatically when the patient removes their mask.
Configuration
Parameter
Description
Pressure Units
Sets the unit in which pressure is displayed
Other options
Button
Description
Therapy/Mask
Launches the Sleep screen, which provides access to mask fit and therapy functions.
Fit
Restore default
Tapping the Restore default settings button opens a confirmation dialog. Once you confirm that you
settings
want to restore default settings, factory default clinical settings will be restored.
Note: Default language, date and time settings will not be restored.
Erase AirMini
Tapping the Erase AirMini data on machine button opens a confirmation dialog. Once you confirm
data on
that you want to erase data, information stored on the AirMini machine will be erased.
machine
Note: Settings, date, time and machine run hours will not be erased.
Erase AirMini
Tapping the Erase AirMini data on phone button opens a confirmation dialog. Once you confirm that
data on phone
you want to erase therapy data, information stored on the phone will be erased.
Upload data
From Machine settings screen, tapping Upload data now opens a dialog box. Once you confirm that you
now
have patient consent, data will be sent to the cloud.
To change a setting:
1. Tap the parameter you want to adjust.
2. Use the spinner (or switches) to choose the required setting.
3. Repeat this process for each parameter to be changed.
4. Tap Apply changes to save your changes and to send the new settings to the machine.
The LED on the machine blinks three times to indicate the settings have been changed.
5. To exit clinical mode, tap Exit.
A confirmation dialog appears.
6. Tap Exit.
Description
Sets the EPR value.
Mode
AutoSet
AutoSet
for Her
Range
CPAP
1 / 2 / 3 cm H
2
Selection
Off / On
Off / On
Selection
cm H
O / hPa
2
O (1 / 2 / 3 hPa)
English
15

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents