It is recommended that the machine is at a minimum distance of 1.1 cm from the body during operation. Not applicable to
masks, air tubing or accessories.
IEC 60601-1 (Edition 3.1) classification
Class II (double insulation), Type BF, Ingress protection IP22.
Operating pressure range
AutoSet, AutoSet For Her, CPAP:
Maximum single fault steady pressure
Machine will shut down in the presence of a single fault if the steady state pressure exceeds 30 cm H
than 6 seconds or 40 cm H
Pneumatic flow path
Pressure accuracy
Maximum static pressure variation at 10 cm H
Tested with full face mask: ±0.5 cm H
Maximum dynamic pressure variation according to ISO 80601-2-70:2015
Machine with full face mask
Pressure
[cm H
O (hPa)]
2
4
8
12
16
20
Flow (maximum) at set pressures
The following are measured accordingly to ISO 80601-2-70:2015 at the end of the specified air tubing:
Pressure
cm H
O (hPa)
2
4
8
12
16
20
O (40 hPa) for more than 1 second.
2
2
0 (0.5 hPa)
2
10 BPM
0.5
0.5
0.5
0.5
0.5
AirMini machine and AirMini air tubing
L/min (including uncertainty)
119
116
112
108
105
4 to 20 cm H
2
1. Flow sensor
2. Blower
3. Pressure sensor
4. AAV (F20 connector only)
5. Vent
6. HumidX (N20, P10 connectors only)
7. Mask
8. Air tubing
9. Machine
10. Inlet filter
O (10 hPa) according to ISO 80601-2-70:2015:
15 BPM
0.7
0.7
0.7
0.7
0.7
O (4 to 20 hPa)
20 BPM
1.0
1.0
1.0
1.0
1.0
O (30 hPa) for more
2
English
27