Endress+Hauser Proline Prowirl R 200 Operating Instructions Manual
Endress+Hauser Proline Prowirl R 200 Operating Instructions Manual

Endress+Hauser Proline Prowirl R 200 Operating Instructions Manual

Hart vortex flowmeter
Hide thumbs Also See for Proline Prowirl R 200:
Table of Contents

Advertisement

Quick Links

BA01688D/06/EN/01.18
71384784
Valid as of version
01.03.zz (Device firmware)
Products
Operating Instructions
Proline Prowirl R 200
HART
Vortex flowmeter
Solutions
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline Prowirl R 200

  • Page 1 Products Solutions Services BA01688D/06/EN/01.18 71384784 Valid as of version 01.03.zz (Device firmware) Operating Instructions Proline Prowirl R 200 HART Vortex flowmeter...
  • Page 2 • The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser Sales Center will supply you with current information and updates to these instructions. Endress+Hauser...
  • Page 3: Table Of Contents

    Proline Prowirl R 200 HART Table of contents Table of contents About this document ....6 Installation ..... . . 22 Document function .
  • Page 4 Table of contents Proline Prowirl R 200 HART 8.3.10 User roles and related access 10.9 Application-specific commissioning ..127 authorization ....63 10.9.1 Steam application .
  • Page 5 13.2 Measuring and test equipment ... 165 13.3 Endress+Hauser services ....165 Repairs ......166 14.1 General notes .
  • Page 6: About This Document

    About this document Proline Prowirl R 200 HART About this document Document function These Operating Instructions contain all the information that is required in various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal.
  • Page 7: Tool Symbols

    Proline Prowirl R 200 HART About this document 1.2.4 Tool symbols Symbol Meaning Flat blade screwdriver Allen key Open-ended wrench 1.2.5 Symbols for certain types of information Symbol Meaning Permitted Procedures, processes or actions that are permitted. Preferred Procedures, processes or actions that are preferred.
  • Page 8: Documentation

    • The W@M Device Viewer : Enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate. For a detailed list of the individual documents along with the documentation code →...
  • Page 9: Basic Safety Instructions

    Proline Prowirl R 200 HART Basic safety instructions Basic safety instructions Requirements for the personnel The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task.
  • Page 10: Workplace Safety

    Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
  • Page 11: It Security

    Proline Prowirl R 200 HART Basic safety instructions IT security We only provide a warranty if the device is installed and used as described in the Operating Instructions. The device is equipped with security mechanisms to protect it against any inadvertent changes to the device settings.
  • Page 12: Product Description

    Product description Proline Prowirl R 200 HART Product description The device consists of a transmitter and a sensor. Two device versions are available: • Compact version – transmitter and sensor form a mechanical unit. • Remote version - transmitter and sensor are mounted in separate locations.
  • Page 13 Proline Prowirl R 200 HART Product description A0034152  2 Versions of pressure measuring unit Order code for "Sensor version", option DA "mass steam" Order code for "Sensor version", option DB "mass gas/liquid" Endress+Hauser...
  • Page 14: Incoming Acceptance And Product

    A0028673 and documents present? • If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center. • Depending on the device version, the CD-ROM might not be part of the delivery! The Technical Documentation is available via the Internet or via the Endress+Hauser Operations App, see the "Product identification"...
  • Page 15: Transmitter Nameplate

    "Supplementary device-dependent documentation" →  8 • The W@M Device Viewer: Enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate. 4.2.1...
  • Page 16: Sensor Nameplate

    Incoming acceptance and product identification Proline Prowirl R 200 HART 4.2.2 Sensor nameplate Order code for "Housing" option B "GT18 two-chamber, 316L, compact" and option K "GT18 two-chamber, 316L, remote" Ser. no.: Size: Qmax(G): Ptest: Materials: Gasket: A0034423  4...
  • Page 17 Proline Prowirl R 200 HART Incoming acceptance and product identification Order code for "Housing" option C "GT20 two-chamber, aluminum, coated, compact" Ser. no.: Size: Qmax(G): Ptest: Materials: Gasket: A0034161  5 Example of a sensor nameplate Nominal diameter of sensor...
  • Page 18 Incoming acceptance and product identification Proline Prowirl R 200 HART Order code for "Housing" option J "GT20 two-chamber, aluminum, coated, remote" Order code: Ser. no.: Ext. ord. cd.: Size: Qmax(G): Ptest: Materials: Gasket: A0034162  6 Example of a sensor nameplate...
  • Page 19: Pressure Measuring Cells Nameplate

    Proline Prowirl R 200 HART Incoming acceptance and product identification 4.2.3 Pressure measuring cells nameplate DCP21 Date: A0034354  7 Example of pressure measuring cell nameplate Manufacturer address Pressure range Maximum permitted pressure Ambient temperature range Serial number or XPD structure...
  • Page 20: Storage And Transport

    Storage and transport Proline Prowirl R 200 HART Storage and transport Storage conditions Observe the following notes for storage: ‣ Store in the original packaging to ensure protection from shock. ‣ Do not remove protective covers or protective caps installed on process connections.
  • Page 21: Measuring Devices With Lifting Lugs

    Proline Prowirl R 200 HART Storage and transport 5.2.2 Measuring devices with lifting lugs CAUTION Special transportation instructions for devices with lifting lugs ‣ Only use the lifting lugs fitted on the device or flanges to transport the device. ‣...
  • Page 22: Installation

    Installation Proline Prowirl R 200 HART Installation Installation conditions 6.1.1 Mounting position Mounting location A0015543 Orientation The direction of the arrow on the sensor nameplate helps you to install the sensor according to the flow direction (direction of medium flow through the piping).
  • Page 23 Proline Prowirl R 200 HART Installation Pressure measuring cell Steam pressure measurement Option DA • With the  transmitter installed at the bottom or at the side • Protection against rising A0034057 heat  • Reduction in temperature to almost ambient...
  • Page 24 Installation Proline Prowirl R 200 HART The following dimensions must be observed to guarantee problem-free access to the device for service purposes: • A =100 mm (3.94 in) • L = L + 150 mm (5.91 in) Inlet and outlet runs To attain the specified level of accuracy of the measuring device, the inlet and outlet runs mentioned below must be maintained at the very minimum.
  • Page 25 Proline Prowirl R 200 HART Installation The flow conditioner is fitted between two pipe flanges and centered by the mounting bolts. Generally this reduces the inlet run needed to 10 × DN with full accuracy. × × × A0019208 Flow conditioner The pressure loss for flow conditioners is calculated as follows: ∆...
  • Page 26: Requirements From Environment And Process

    ‣ If operating outdoors: Avoid direct sunlight, particularly in warm climatic regions. You can order a weather protection cover from Endress+Hauser. →  169. Thermal insulation For optimum temperature measurement and mass calculation, heat transfer at the sensor must be avoided for some fluids. This can be ensured by installing thermal insulation. A wide range of materials can be used for the required insulation.
  • Page 27: Special Mounting Instructions

    Proline Prowirl R 200 HART Installation A0019212 Maximum insulation height ‣ When insulating, ensure that a sufficiently large area of the housing support remains exposed. The uncovered part serves as a radiator and protects the electronics from overheating and excessive cooling.
  • Page 28: Mounting The Measuring Device

    Installation Proline Prowirl R 200 HART A0019209  9 Layout for delta heat measurement of saturated steam and water Measuring device Temperature sensor Heat exchanger Heat flow Protective cover Observe the following minimum head clearance: 222 mm (8.74 in) For information on the weather protection cover, see →  169 Mounting the measuring device 6.2.1...
  • Page 29: Mounting The Pressure Measuring Unit

    Proline Prowirl R 200 HART Installation 1. Ensure that the direction of the arrow on the sensor matches the flow direction of the medium. 2. To ensure compliance with device specifications, install the measuring device between the pipe flanges in a way that it is centered in the measurement section.
  • Page 30 Installation Proline Prowirl R 200 HART Mounting the pressure measuring unit 6 mm A0035442 Siphon Pressure measuring cell NOTICE Damage to seal! The seal is made of expanded graphite. It can therefore be used only once. If a coupling is released, a new seal must be installed.
  • Page 31: Mounting The Transmitter Of The Remote Version

    Proline Prowirl R 200 HART Installation Connecting the pressure measuring unit A0035443 Device plug 6. Insert the plug for electrical connection of the pressure measuring cell and screw into place. 6.2.5 Mounting the transmitter of the remote version CAUTION Ambient temperature too high! Danger of electronics overheating and housing deformation.
  • Page 32: Turning The Transmitter Housing

    Installation Proline Prowirl R 200 HART Wall mounting 19 (0.6) 80 (3.15) A0033484  10 mm (in) Post mounting 20 70 … ( 0.79 to 2.75) SW 13 A0033486  11 mm (in) 6.2.6 Turning the transmitter housing To provide easier access to the connection compartment or display module, the transmitter housing can be turned.
  • Page 33: Turning The Display Module

    Proline Prowirl R 200 HART Installation max. 350° 8 mm 8 mm A0032242 1. Release the fixing screw. 2. Turn the housing to the desired position. 3. Firmly tighten the securing screw. 6.2.7 Turning the display module The display module can be turned to optimize display readability and operability.
  • Page 34 Installation Proline Prowirl R 200 HART Has the correct orientation for the sensor been selected →  22? • According to sensor type  • According to medium temperature • According to medium properties (outgassing, with entrained solids) Does the arrow on the sensor nameplate match the direction of flow of the fluid through the ...
  • Page 35: Electrical Connection

    Proline Prowirl R 200 HART Electrical connection Electrical connection Connection conditions 7.1.1 Required tools • For cable entries: Use corresponding tools • For securing clamp: Allen key 3 mm • Wire stripper • When using stranded cables: Crimper for wire end ferrule •...
  • Page 36: Connecting Cable For Remote Version

    Electrical connection Proline Prowirl R 200 HART 7.1.3 Connecting cable for remote version Connecting cable (standard) Standard cable 2 × 2 × 0.5 mm (22 AWG) PVC cable with common shield (2 pairs, pair- stranded) Flame resistance According to DIN EN 60332-1-2...
  • Page 37: Terminal Assignment

    Proline Prowirl R 200 HART Electrical connection 7.1.4 Terminal assignment Transmitter 4-20 mA HART connection version with additional inputs and outputs – – – – – – A0033475 A0033475 Maximum number of terminals Maximum number of terminals for order code for Terminals 1 to 6: "Accessory mounted", option NA "Overvoltage...
  • Page 38 Electrical connection Proline Prowirl R 200 HART Connecting cable for remote version Transmitter and sensor connection housing In the case of the remote version, the sensor and transmitter are mounted separately from on another and connected by a connecting cable. Connection is performed via the sensor connection housing and the transmitter housing.
  • Page 39: Requirements For The Supply Unit

    Proline Prowirl R 200 HART Electrical connection – – BK YE BU – – A0034571  13 Terminals for connection compartment in the transmitter wall holder and the sensor connection housing Terminals for connecting cable Grounding via the cable strain relief...
  • Page 40 Electrical connection Proline Prowirl R 200 HART Increase in minimum terminal voltage Increase in minimum Local operation terminal voltage Order code for "Display; Operation", option C: + DC 1 V Local operation SD02 Order code for "Display; Operation", option E:...
  • Page 41: Preparing The Measuring Device

    Proline Prowirl R 200 HART Electrical connection 7.1.6 Preparing the measuring device Carry out the steps in the following order: 1. Mount the sensor and transmitter. 2. Connection housing, sensor: Connect connecting cable. 3. Transmitter: Connect connecting cable. 4. Transmitter: Connect signal cable and cable for supply voltage.
  • Page 42 Electrical connection Proline Prowirl R 200 HART Connection via terminals 3 mm 20 mm 10 (0.4) mm (in) A0032239 1. Loosen the securing clamp of the connection compartment cover. 2. Unscrew the connection compartment cover. 3. Push the cable through the cable entry . To ensure tight sealing, do not remove the sealing ring from the cable entry.
  • Page 43: Connecting The Remote Version

    Proline Prowirl R 200 HART Electrical connection ‣ To remove a cable from the terminal, use a flat-blade screwdriver to push the slot between the two terminal holes while simultaneously pulling the cable end out of the terminal. 7.2.2 Connecting the remote version...
  • Page 44 Electrical connection Proline Prowirl R 200 HART A0034171  15 Sample graphic Connecting cable (standard, reinforced) 3. Guide the connecting cable through the cable entry and into the connection housing (if using a connecting cable without an M12 device plug, use the shorter stripped end of the connecting cable).
  • Page 45 Proline Prowirl R 200 HART Electrical connection Connecting the transmitter Connecting transmitter via plug A0034172 ‣ Connect the plug. Connecting transmitter via terminals 3 mm A0034173 1. Loosen the securing clamp of the electronics compartment cover. 2. Unscrew the electronics compartment cover.
  • Page 46 Electrical connection Proline Prowirl R 200 HART ~15° A0034175  16 Sample graphic 6. Turn the transmitter housing to the right until it reaches the marking. NOTICE The connection board of the wall housing is connected to the electronics board of the transmitter via a signal cable! ‣...
  • Page 47: Connecting The Connecting Cable For The Pressure Measuring Cell

    Proline Prowirl R 200 HART Electrical connection Connecting cable (standard, reinforced) 8. Disconnect the signal cable from the connection board of the wall housing . by pressing in the locking clip on the connector. Remove the transmitter housing. 9. Guide the connecting cable through the cable entry and into the connection housing (if using a connecting cable without an M12 device plug, use the shorter stripped end of the connecting cable).
  • Page 48: Ensuring Potential Equalization

    Electrical connection Proline Prowirl R 200 HART 7.2.4 Ensuring potential equalization Requirements Please consider the following to ensure correct measurement: • Same electrical potential for the medium and sensor • Remote version: same electrical potential for the sensor and transmitter •...
  • Page 49 Proline Prowirl R 200 HART Electrical connection Have the screws for the cable strain relief been tightened using the correct torque→  43?  Has the M12 plug of the connecting cable been correctly connected to the pressure measuring ...
  • Page 50: Operation Options

    Operation options Proline Prowirl R 200 HART Operation options Overview of operation options A0032226 Local operation via display module Computer with operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) Field Xpert SFX350 or SFX370 Field Communicator 475 Control system (e.g. PLC)
  • Page 51: Structure And Function Of The Operating Menu

    Proline Prowirl R 200 HART Operation options Structure and function of the operating menu 8.2.1 Structure of the operating menu For an overview of the operating menu for experts: "Description of Device Parameters" document supplied with the device Operating menu for operators and maintenances...
  • Page 52: Operating Philosophy

    Operation options Proline Prowirl R 200 HART 8.2.2 Operating philosophy The individual parts of the operating menu are assigned to certain user roles (operator, maintenance etc.). Each user role contains typical tasks within the device lifecycle. Menu/parameter User role and tasks...
  • Page 53: Access To The Operating Menu Via The Local Display

    Proline Prowirl R 200 HART Operation options Access to the operating menu via the local display 8.3.1 Operational display X X X X X X X 1120.50 A0029346 Operational display Device tag→  75 Status area Display area for measured values (4-line) Operating elements→...
  • Page 54: Navigation View

    Operation options Proline Prowirl R 200 HART Totalizer  The measurement channel number indicates which of the three totalizers is displayed. Output  The measurement channel number indicates which of the two current outputs is displayed. Measurement channel numbers Symbol...
  • Page 55 Proline Prowirl R 200 HART Operation options For more information about the icons in the menu, refer to the "Display area" section →  55 Status area The following appears in the status area of the navigation view in the top right corner: •...
  • Page 56: Editing View

    Operation options Proline Prowirl R 200 HART Wizard operation Symbol Meaning Switches to the previous parameter. Confirms the parameter value and switches to the next parameter. Opens the editing view of the parameter. 8.3.3 Editing view Numeric editor Text editor...
  • Page 57 Proline Prowirl R 200 HART Operation options Text editor Symbol Meaning Toggle • Between upper-case and lower-case letters • For entering numbers • For entering special characters Selection of letters from A to Z. ABC _ … Selection of letters from a to z.
  • Page 58: Operating Elements

    Operation options Proline Prowirl R 200 HART 8.3.4 Operating elements Operating key(s) Meaning Minus key In a menu, submenu Moves the selection bar upwards in a choose list. With a Wizard Confirms the parameter value and goes to the previous parameter.
  • Page 59 Proline Prowirl R 200 HART Operation options • Setup • Configuration backup display • Simulation Calling up and closing the context menu The user is in the operational display. 1. Press  for 2 s.  The context menu opens.
  • Page 60: Navigating And Selecting From List

    Operation options Proline Prowirl R 200 HART 8.3.6 Navigating and selecting from list Different operating elements are used to navigate through the operating menu. The navigation path is displayed on the left in the header. Icons are displayed in front of the individual menus.
  • Page 61: Calling Up Help Text

    Proline Prowirl R 200 HART Operation options The direct access code consists of a 5-digit number (at maximum) and the channel number, which identifies the channel of a process variable: e.g. 00914-2. In the navigation view, this appears on the right-hand side in the header of the selected parameter.
  • Page 62: Changing The Parameters

    Operation options Proline Prowirl R 200 HART 8.3.9 Changing the parameters For a description of the editing view - consisting of the text editor and numeric editor - with symbols →  56, for a description of the operating elements →  58 Example: Changing the tag name in the "Tag description"...
  • Page 63: User Roles And Related Access Authorization

    Proline Prowirl R 200 HART Operation options Ent. access code Invalid or out of range input value Min:0 Max:9999 A0014049-EN 8.3.10 User roles and related access authorization The two user roles "Operator" and "Maintenance" have different write access to the parameters if the customer defines a user-specific access code.
  • Page 64: Enabling And Disabling The Keypad

    Operation options Proline Prowirl R 200 HART 8.3.12 Enabling and disabling the keypad lock The keypad lock makes it possible to block access to the entire operating menu via local operation. As a result, it is no longer possible to navigate through the operating menu or change the values of individual parameters.
  • Page 65: Field Xpert Sfx350, Sfx370

    VIATOR Bluetooth modem with connecting cable Transmitter Via service interface (CDI) A0034056 Service interface (CDI = Endress+Hauser Common Data Interface) of the measuring device Commubox FXA291 Computer with FieldCare operating tool with COM DTM CDI Communication FXA291 8.4.2 Field Xpert SFX350, SFX370...
  • Page 66: Fieldcare

    FieldCare Function scope FDT-based plant asset management tool from Endress+Hauser. It can configure all smart field devices in a system and helps you manage them. By using the status information, it is also a simple but effective way of checking their status and condition.
  • Page 67: Devicecare

    DeviceCare Function scope Tool to connect and configure Endress+Hauser field devices. The fastest way to configure Endress+Hauser field devices is with the dedicated "DeviceCare" tool. Together with the device type managers (DTMs) it presents a convenient, comprehensive solution. For details, see Innovation Brochure IN01047S Source for device description files See information →...
  • Page 68: Simatic Pdm

    Operation options Proline Prowirl R 200 HART Source for device description files See data →  69 8.4.6 SIMATIC PDM Function scope SIMATIC PDM is a standardized, manufacturer-independent program from Siemens for the operation, configuration, maintenance and diagnosis of intelligent field devices via HART protocol.
  • Page 69: System Integration

    Proline Prowirl R 200 HART System integration System integration Overview of device description files 9.1.1 Current version data for the device Firmware version 01.03.00 • On the title page of the Operating Instructions • On the transmitter nameplate • Firmware version parameter Diagnostics →...
  • Page 70 System integration Proline Prowirl R 200 HART Dynamic variables Measured variables (HART device variables) Primary dynamic variable (PV) Volume flow Secondary dynamic variable (SV) Temperature Tertiary dynamic variable (TV) Totalizer 1 Quaternary dynamic variable (QV) Totalizer 2 The assignment of the measured variables to the dynamic variables can be modified and...
  • Page 71: Other Settings

    Proline Prowirl R 200 HART System integration • 7 = total mass flow • 8 = energy flow • 9 = heat flow difference • 17 = pressure Other settings Burst mode functionality in accordance with HART 7 Specification: Navigation "Expert"...
  • Page 72 System integration Proline Prowirl R 200 HART Parameter overview with brief description Parameter Description Selection / User entry Factory setting Burst mode 1 to n Activate the HART burst mode for burst • Off message X. • On Burst command 1 to n Select the HART command that is sent to the •...
  • Page 73 Proline Prowirl R 200 HART System integration Parameter Description Selection / User entry Factory setting Min. update period Enter the minimum time span between two Positive integer 1 000 ms burst commands of burst message X. Max. update period Enter the maximum time span between two...
  • Page 74: Commissioning

    Commissioning Proline Prowirl R 200 HART Commissioning 10.1 Function check Before commissioning the measuring device: ‣ Make sure that the post-installation and post-connection checks have been performed. • "Post-installation check" checklist→  33 • "Post-connection check" checklist →  48 10.2...
  • Page 75: Configuring The Measuring Device

    Proline Prowirl R 200 HART Commissioning 10.4 Configuring the measuring device • The Setup menu with its guided wizards contains all the parameters needed for standard operation. • Navigation to the Setup menu X X X X X X X 20.50...
  • Page 76: Setting The System Units

    Commissioning Proline Prowirl R 200 HART XXXXXXXXX A0029422  24 Header of the operational display with tag name Tag name Enter the tag name in the "FieldCare" operating tool →  67 Navigation "Setup" menu → Device tag Parameter overview with brief description...
  • Page 77 Proline Prowirl R 200 HART Commissioning Energy unit →  78 Calorific value unit →  78 Calorific value unit →  79 Velocity unit →  79 →  79 Density unit Specific volume unit →  79 Dynamic viscosity unit →...
  • Page 78 Commissioning Proline Prowirl R 200 HART Parameter Prerequisite Description Selection Factory setting Pressure unit With order code for "Sensor Select process pressure unit. Unit choose list Country-specific: version": • bar Result • option "Mass (integrated • psi The unit is taken from: temperature measurement)"...
  • Page 79 Proline Prowirl R 200 HART Commissioning Parameter Prerequisite Description Selection Factory setting Calorific value unit (Mass) The following conditions are Select calorific value unit. Unit choose list Country-specific: met: • kJ/kg • Order code for "Sensor • Btu/lb version", – Option "Mass (integrated temperature measurement)"...
  • Page 80: Selecting And Setting The Medium

    Commissioning Proline Prowirl R 200 HART 10.4.3 Selecting and setting the medium The Medium selection wizard systematically guides the user through all the parameters that must be configured in order to select and set the medium. Navigation "Setup" menu → Medium selection ‣...
  • Page 81 Proline Prowirl R 200 HART Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Gas type The following conditions are Select measured gas type. • Hydrogen H2 Methane CH4 met: • Helium He • In the Select medium • Neon Ne parameter, the Gas option is •...
  • Page 82: Configuring The Current Input

    Commissioning Proline Prowirl R 200 HART Parameter Prerequisite Description Selection / User Factory setting entry Fixed process pressure The following conditions are Enter fixed value for process 0 to 250 bar abs. 0 bar abs. met: pressure. • Order code for "Sensor Dependency version",...
  • Page 83 Proline Prowirl R 200 HART Commissioning Navigation "Setup" menu → Current input ‣ Current input →  83 External value →  83 Atmospheric pressure Current span →  83 4 mA value →  83 20 mA value →  83 Failure mode →...
  • Page 84: Configuring The Current Output

    Commissioning Proline Prowirl R 200 HART 10.4.5 Configuring the current output The Current output wizard guides you systematically through all the parameters that have to be set for configuring the current output. Navigation "Setup" menu → Current output 1 to n ‣...
  • Page 85: Configuring The Pulse/Frequency

    Proline Prowirl R 200 HART Commissioning Parameter Prerequisite Description Selection / User Factory setting entry 20 mA value One of the following options is Enter 20 mA value. Signed floating-point Depends on country selected in the Current span number and nominal parameter (→  84): diameter •...
  • Page 86 Commissioning Proline Prowirl R 200 HART Configuring the pulse output Navigation "Setup" menu → Pulse/frequency/switch output Pulse/frequency/switch output Assign pulse output 1 →  86 Value per pulse →  86 Pulse width →  86 →  87 Failure mode →...
  • Page 87 Proline Prowirl R 200 HART Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Failure mode In the Operating mode Define output behavior in • Actual value No pulses parameter, the Pulse option is alarm condition. • No pulses selected, and one of the...
  • Page 88 Commissioning Proline Prowirl R 200 HART Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Assign frequency output The Frequency option is Select process variable for • Off selected in the Operating frequency output. • Volume flow mode parameter (→...
  • Page 89 Proline Prowirl R 200 HART Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Measuring value at minimum In the Operating mode Enter measured value for Signed floating-point Depends on country frequency parameter, the Frequency minmum frequency. number and nominal option is selected, and one of...
  • Page 90 Commissioning Proline Prowirl R 200 HART Parameter Prerequisite Description Selection / User Factory setting entry Failure frequency In the Operating mode Enter frequency output value 0.0 to 1 250.0 Hz 0.0 Hz parameter (→  85), the in alarm condition. Frequency option is selected,...
  • Page 91 Proline Prowirl R 200 HART Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Switch output function The Switch option is selected Select function for switch • Off in the Operating mode output. • On parameter.
  • Page 92: Configuring The Local Display

    Commissioning Proline Prowirl R 200 HART Parameter Prerequisite Description Selection / User Factory setting entry Switch-on delay • The Switch option is Define delay for the switch-on 0.0 to 100.0 s 0.0 s selected in the Operating of status output. mode parameter.
  • Page 93 Proline Prowirl R 200 HART Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Format display A local display is provided. Select how measured values • 1 value, max. size 1 value, max. size are shown on the display.
  • Page 94: Configuring The Output

    Commissioning Proline Prowirl R 200 HART 10.4.8 Configuring the output conditioning The Output conditioning wizard guides you systematically through all the parameters that have to be set for configuring the output conditioning. Navigation "Setup" menu → Output conditioning ‣ Output conditioning →...
  • Page 95 Proline Prowirl R 200 HART Commissioning Navigation "Setup" menu → Low flow cut off ‣ Low flow cut off →  95 Sensitivity →  95 Turn down Parameter overview with brief description Parameter Description User entry Factory setting Sensitivity...
  • Page 96: Advanced Settings

    Commissioning Proline Prowirl R 200 HART 10.5 Advanced settings The Advanced setup submenu together with its submenus contains parameters for specific settings. Navigation to the "Advanced setup" submenu X X X X X X X 20.50 0104-1 Main menu Language...
  • Page 97: Setting The Medium Properties

    Proline Prowirl R 200 HART Commissioning ‣ SIL confirmation ‣ Deactivate SIL ‣ Display →  116 ‣ Heartbeat setup ‣ Configuration backup display →  118 ‣ Administration →  120 10.5.1 Setting the medium properties In the Medium properties submenu the reference values for the measuring application can be set.
  • Page 98 Commissioning Proline Prowirl R 200 HART Dynamic viscosity →  100 ‣ Gas composition →  100 Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Enthalpy type The following conditions are Define which kind of enthalpy •...
  • Page 99 Proline Prowirl R 200 HART Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Reference pressure The following conditions are Enter reference pressure for 0 to 250 bar 1.01325 bar met: the calulation of the reference • Order code for "Sensor density.
  • Page 100 Commissioning Proline Prowirl R 200 HART Parameter Prerequisite Description Selection / User Factory setting entry Calorific value The following conditions are Enter gross calorific value to Positive floating- 50 000 kJ/kg met: calculate the energy flow. point number • Selected medium: – In the Select gas type...
  • Page 101 Proline Prowirl R 200 HART Commissioning Navigation "Setup" menu → Advanced setup → Medium properties → Gas composition ‣ Gas composition Gas mixture →  103 Mol% Ar →  103 Mol% C2H3Cl →  103 Mol% C2H4 →  104 Mol% C2H6 →...
  • Page 102 Commissioning Proline Prowirl R 200 HART Mol% n-C6H14 →  108 Mol% n-C7H16 →  108 Mol% n-C8H18 →  109 Mol% n-C9H20 →  109 →  109 Mol% Ne Mol% NH3 →  109 Mol% O2 →  109 Mol% SO2 →...
  • Page 103 Proline Prowirl R 200 HART Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Gas mixture The following conditions are Select measured gas mixture. • Hydrogen H2 Methane CH4 met: • Helium He • In the Select medium •...
  • Page 104 Commissioning Proline Prowirl R 200 HART Parameter Prerequisite Description Selection / User Factory setting entry Mol% C2H4 The following conditions are Enter amount of substance for 0 to 100 % met: the gas mixture. • In the Select medium parameter, the Gas option is selected.
  • Page 105 Proline Prowirl R 200 HART Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Mol% Cl2 The following conditions are Enter amount of substance for 0 to 100 % met: the gas mixture. • In the Select medium parameter, the Gas option is selected.
  • Page 106 Commissioning Proline Prowirl R 200 HART Parameter Prerequisite Description Selection / User Factory setting entry Mol% H2O The following conditions are Enter amount of substance for 0 to 100 % met: the gas mixture. • In the Select medium parameter, the Gas option is selected.
  • Page 107 Proline Prowirl R 200 HART Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Mol% i-C4H10 The following conditions are Enter amount of substance for 0 to 100 % met: the gas mixture. • In the Select medium parameter, the Gas option is selected.
  • Page 108 Commissioning Proline Prowirl R 200 HART Parameter Prerequisite Description Selection / User Factory setting entry Mol% n-C4H10 The following conditions are Enter amount of substance for 0 to 100 % met: the gas mixture. • In the Select medium parameter, the Gas option is selected.
  • Page 109 Proline Prowirl R 200 HART Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Mol% n-C8H18 The following conditions are Enter amount of substance for 0 to 100 % met: the gas mixture. • In the Select medium parameter, the Gas option is selected.
  • Page 110: Performing External Compensation

    Commissioning Proline Prowirl R 200 HART Parameter Prerequisite Description Selection / User Factory setting entry Mol% SO2 The following conditions are Enter amount of substance for 0 to 100 % met: the gas mixture. • In the Select medium parameter, the Gas option is selected.
  • Page 111 Proline Prowirl R 200 HART Commissioning 2nd temperature delta heat →  112 Fixed process pressure →  112 Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry External value With order code for "Sensor Assign variable from external •...
  • Page 112: Carrying Out A Sensor Adjustment

    Commissioning Proline Prowirl R 200 HART Parameter Prerequisite Description Selection / User Factory setting entry 2nd temperature delta heat The 2nd temperature delta Enter 2nd temperature value –200 to 450 °C 20 °C heat parameter is visible. to calculate the delta heat.
  • Page 113 Proline Prowirl R 200 HART Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Inlet configuration The inlet run correction Select inlet configuration. • Off feature: • Single elbow • Is a standard feature and • Double elbow can only be used in Prowirl F •...
  • Page 114: Configuring The Totalizer

    Commissioning Proline Prowirl R 200 HART Example Influence of the diameter mismatch without using the correction function: • Mating pipe DN 100 (4"), schedule 80 • Device flange DN 100 (4"), schedule 40 • This installation position results in a diameter mismatch of 5 mm (0.2 in). If the correction function is not used, an additional measuring uncertainty of approx.
  • Page 115 Proline Prowirl R 200 HART Commissioning Parameter Prerequisite Description Selection Factory setting Totalizer operation mode – Select totalizer calculation • Net flow total Net flow total mode. • Forward flow total • Reverse flow total Failure mode One of the following options is Define totalizer behavior in •...
  • Page 116: Carrying Out Additional Display Configurations

    Commissioning Proline Prowirl R 200 HART 10.5.5 Carrying out additional display configurations In the Display submenu you can set all the parameters associated with the configuration of the local display. Navigation "Setup" menu → Advanced setup → Display ‣ Display →...
  • Page 117 Proline Prowirl R 200 HART Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Format display A local display is provided. Select how measured values • 1 value, max. size 1 value, max. size are shown on the display.
  • Page 118: Configuration Management

    Commissioning Proline Prowirl R 200 HART Parameter Prerequisite Description Selection / User Factory setting entry Decimal places 3 A measured value is specified Select the number of decimal • x x.xx in the Value 3 display places for the display value. • x.x parameter.
  • Page 119 Proline Prowirl R 200 HART Commissioning You can do so using the Configuration management parameter and the related options found in the Configuration backup display submenu. Navigation "Setup" menu → Advanced setup → Configuration backup display ‣ Configuration backup display Operating time →...
  • Page 120: Using Parameters For Device Administration

    Commissioning Proline Prowirl R 200 HART Options Description Duplicate The transmitter configuration from another device is duplicated to the device using the display module. Clear backup data The backup copy of the device configuration is deleted from the display module of the device.
  • Page 121: Function Scope Of The "Configuration Management" Parameter

    Proline Prowirl R 200 HART Commissioning Navigation "Setup" menu → Advanced setup → Configuration backup display ‣ Configuration backup display →  119 Operating time →  119 Last backup Configuration management →  119 Comparison result →  119...
  • Page 122: Simulation

    Commissioning Proline Prowirl R 200 HART Options Description Duplicate The transmitter configuration from another device is duplicated to the device using the display module. Clear backup data The backup copy of the device configuration is deleted from the display module of the device.
  • Page 123 Proline Prowirl R 200 HART Commissioning Diagnostic event category →  124 Diagnostic event simulation →  124 Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Assign simulation process variable – Select a process variable for •...
  • Page 124: Protecting Settings From Unauthorized

    Commissioning Proline Prowirl R 200 HART Parameter Prerequisite Description Selection / User Factory setting entry Pulse output simulation In the Operating mode Set and switch off the pulse • Off parameter, the Pulse option is output simulation. • Fixed value selected. • Down-counting ...
  • Page 125: Write Protection Via Write Protection

    Proline Prowirl R 200 HART Commissioning 3. Enter the access code again in the to confirm the code.  The -symbol appears in front of all write-protected parameters. The device automatically locks the write-protected parameters again if a key is not pressed for 10 minutes in the navigation and editing view.
  • Page 126 Commissioning Proline Prowirl R 200 HART XXXXXXXXXXXXX 3 mm A0032230 1. Loosen the securing clamp. 2. Unscrew the electronics compartment cover. 3. Pull out the display module with a gentle rotational movement. To make it easier to access the write protection switch, attach the display module to the edge of the electronics compartment.
  • Page 127: Application-Specific Commissioning

    Proline Prowirl R 200 HART Commissioning 4. Setting the write protection switch (WP) on the main electronics module to the ON position enables hardware write protection. Setting the write protection switch (WP) on the main electronics module to the OFF position (factory setting) disables hardware write protection.
  • Page 128: Gas Applications

    Commissioning Proline Prowirl R 200 HART Select medium Navigation: Setup → Medium selection 1. Call up the Medium selection wizard. 2. In the Select medium parameter, select the Liquid option. 3. In the Select liquid type parameter, select the User-specific liquid option.
  • Page 129 Proline Prowirl R 200 HART Commissioning 6. In the Reference combustion temperature parameter, enter the reference combustion temperature of the fluid. Configuring the current output 8. Configure the current output for the "energy flow" process variable→  84. Configuring optional fluid properties for output of corrected volume flow Navigation: Setup →...
  • Page 130 8. In the Reference temperature parameter (→  99) enter the temperate for calculating the reference density. Endress+Hauser recommends the use of active pressure compensation. This fully rules out the risk of measured errors due to pressure variations and incorrect entries .
  • Page 131 12. In the Relative density parameter, enter the relative density of the natural gas. Endress+Hauser recommends the use of active pressure compensation. This fully rules out the risk of measured errors due to pressure variations and incorrect entries .
  • Page 132: Calculation Of The Measured

    Commissioning Proline Prowirl R 200 HART 10.9.4 Calculation of the measured variables A flow computer can be found in the electronics of the measuring device with order code for "Sensor version", option "mass (integrated temperature measurement)" and option "mass (integrated pressure/temperature measurement)". This computer can calculate the...
  • Page 133 Proline Prowirl R 200 HART Commissioning Energy flow Medium Fluid Standards Explanation Heat/energy option – For fixed process IAPWS- pressure or if the Steam IF97/ASME pressure is read in via current input/HART Single gas • Contains GPA 2172 • For fixed process...
  • Page 134 Commissioning Proline Prowirl R 200 HART Steam is calculated based on the following factors: • Fully compensated calculation of density using the "pressure" and "temperature" measured variables • Calculation based on overheated steam until saturation point is reached Configuration of diagnostic behavior of the diagnostic message S871 Near steam saturation limitAssign behavior of diagnostic no.
  • Page 135 Proline Prowirl R 200 HART Commissioning Ethane Propane Butane Ethylene (ethene) Vinyl chloride Mixtures of up to 8 components of these gases The energy flow is calculated as per ISO 6976 (contains GPA 2172) or AGA5 - in relation to the net calorific value or gross calorific value .
  • Page 136: Operation

    Operation Proline Prowirl R 200 HART Operation 11.1 Reading the device locking status Device active write protection: Locking status parameter Operation → Locking status Function scope of the "Locking status" parameter Options Description None The access status displayed in the Access status display parameter applies →...
  • Page 137: Process Variables

    Proline Prowirl R 200 HART Operation 11.4.1 Process variables The Process variables submenu contains all the parameters needed to display the current measured values for each process variable. Navigation "Diagnostics" menu → Measured values → Process variables ‣ Process variables →...
  • Page 138 Operation Proline Prowirl R 200 HART Parameter Prerequisite Description User interface Mass flow – Displays the mass flow currently Signed floating-point calculated. number Dependency The unit is taken from the Mass flow unit parameter (→  77). Flow velocity –...
  • Page 139: Totalizer" Submenu

    Proline Prowirl R 200 HART Operation Parameter Prerequisite Description User interface Specific volume With order code for "Sensor version": Displays the current value for the Positive floating-point • Option "Mass (integrated specific volume. number temperature measurement)" Dependency The unit is taken from the Specific •...
  • Page 140: Input Values

    Operation Proline Prowirl R 200 HART Parameter overview with brief description Parameter Prerequisite Description User interface Totalizer value 1 to n One of the following options is selected Displays the current totalizer counter Signed floating-point in the Assign process variable value.
  • Page 141: Adapting The Measuring Device To The Process

    Proline Prowirl R 200 HART Operation Navigation "Diagnostics" menu → Measured values → Output values ‣ Output values Output current 1 →  141 Measured current 1 Terminal voltage 1 →  141 Output current 2 Pulse output →  141 Output frequency →...
  • Page 142: Function Scope Of The "Control Totalizer" Parameter

    Operation Proline Prowirl R 200 HART Navigation "Operation" menu → Totalizer handling ‣ Totalizer handling →  142 Control Totalizer 1 to n →  142 Preset value 1 to n Reset all totalizers →  142 Parameter overview with brief description...
  • Page 143: Function Scope Of The "Reset All Totalizers" Parameter

    Proline Prowirl R 200 HART Operation 11.6.2 Function scope of the "Reset all totalizers" parameter Options Description Cancel No action is executed and the user exits the parameter. Reset + totalize Resets all totalizers to 0 and restarts the totaling process. This deletes all the flow values previously totalized.
  • Page 144 Operation Proline Prowirl R 200 HART ‣ Display channel 1 ‣ Display channel 2 ‣ Display channel 3 ‣ Display channel 4 Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Assign channel 1 The Extended HistoROM Assign a process variable to •...
  • Page 145 Proline Prowirl R 200 HART Operation Parameter Prerequisite Description Selection / User Factory setting entry Assign channel 4 The Extended HistoROM Assign process variable to Picklist, see Assign application package is logging channel. channel 1 parameter available. (→  144)  The software options...
  • Page 146: Diagnostics And Troubleshooting

    Diagnostics and troubleshooting Proline Prowirl R 200 HART Diagnostics and troubleshooting 12.1 General troubleshooting For local display Error Possible causes Solution Local display dark and no output Supply voltage does not match the Apply the correct supply voltage signals value indicated on the nameplate.
  • Page 147 Proline Prowirl R 200 HART Diagnostics and troubleshooting Error Possible causes Solution Device shows correct value on local Configuration error Check and correct the parameter display, but signal output is configuration. incorrect, though in the valid range. Device measures incorrectly.
  • Page 148: Diagnostic Information On Local Display

    Diagnostics and troubleshooting Proline Prowirl R 200 HART 12.2 Diagnostic information on local display 12.2.1 Diagnostic message Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display. Operational display in alarm condition...
  • Page 149 Proline Prowirl R 200 HART Diagnostics and troubleshooting Diagnostic behavior Symbol Meaning Alarm • Measurement is interrupted. • Signal outputs and totalizers assume the defined alarm condition. • A diagnostic message is generated. • For local display with touch control: the background lighting changes to red.
  • Page 150: Calling Up Remedial Measures

    Diagnostics and troubleshooting Proline Prowirl R 200 HART 12.2.2 Calling up remedial measures X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage Diagnostics 2...
  • Page 151 Proline Prowirl R 200 HART Diagnostics and troubleshooting Xxxxxx/…/…/ Device name: Xxxxxxx Mass flow: kg/h 12.34 Device tag: Xxxxxxx Volume flow: 12.34 m /h ³ Status signal: Function check (C) Xxxxxx Diagnostics 1: C485 Simu... Remedy information: Deactivate... Failure (F)
  • Page 152: Calling Up Remedy Information

    Diagnostics and troubleshooting Proline Prowirl R 200 HART Diagnostic information Diagnostic code Diagnostic Diagnostic Status signal Short text behavior number ↓ ↓ ↓ Example Curr.output 1 A0013962 A0013958 NAMUR 3-digit number NE 107 12.3.2 Calling up remedy information Remedy information is provided for every diagnostic event to ensure that problems can be rectified quickly: •...
  • Page 153: Adapting The Status Signal

    Proline Prowirl R 200 HART Diagnostics and troubleshooting Options Description Logbook entry only The device continues to measure. The diagnostic message is displayed only in the Event logbook submenu (Event list submenu) and is not displayed in alternation with the operational display.
  • Page 154 Diagnostics and troubleshooting Proline Prowirl R 200 HART Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Sensor connection 1. Check plug connections Alarm defective 2. Change pre-amplifier 3. Change DSC sensor Data storage 1.
  • Page 155 Proline Prowirl R 200 HART Diagnostics and troubleshooting Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Memory content 1. Transfer data or reset device Alarm 2. Contact service Device verification Device verification active, please...
  • Page 156 Diagnostics and troubleshooting Proline Prowirl R 200 HART Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Diagnostic event Deactivate simulation Warning simulation Flow computer Check input value (pressure, Warning configuration incorrect temperature) Flow computer 1.
  • Page 157: Operating Conditions For Displaying The Following Diagnostics

    Proline Prowirl R 200 HART Diagnostics and troubleshooting Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Vibration detected Check installation Warning Vibration exceeded Check installation Alarm Signal quality bad 1. Check process conditions: wet...
  • Page 158: Diagnostic List

    Diagnostics and troubleshooting Proline Prowirl R 200 HART Navigation "Diagnostics" menu  Diagnostics →  158 Actual diagnostics →  158 Previous diagnostics Operating time from restart →  158 Operating time →  158 Parameter overview with brief description...
  • Page 159: Event Logbook

    Proline Prowirl R 200 HART Diagnostics and troubleshooting / ../Diagnose list Diagnostics F273 Main electronic Diagnostics 2 Diagnostics 3 A0014006-EN  27 Taking the example of the local display To call up the measures to rectify a diagnostic event: • Via local display →  150 •...
  • Page 160: Overview Of Information Events

    Diagnostics and troubleshooting Proline Prowirl R 200 HART Navigation path Diagnostics → Event logbook → Filter options Filter categories • All • Failure (F) • Function check (C) • Out of specification (S) • Maintenance required (M) • Information (I) 12.8.3...
  • Page 161: Resetting The Measuring Device

    Proline Prowirl R 200 HART Diagnostics and troubleshooting Info number Info name I1514 Upload started I1515 Upload finished I1552 Failed: Main electronic verification I1553 Failed: Pre-amplifier verification I1554 Safety sequence started I1555 Safety sequence confirmed I1556 Safety mode off 12.9 Resetting the measuring device Using theDevice reset parameter (→...
  • Page 162 Diagnostics and troubleshooting Proline Prowirl R 200 HART Extended order code 2 →  162 Extended order code 3 →  162 ENP version →  162 Device revision →  163 →  163 Device ID Device type →  163 Manufacturer ID →...
  • Page 163: 12.11 Firmware History

    "Manufacturer' s information" document. The manufacturer' s information is available: • In the Download Area of the Endress+Hauser web site: www.endress.com → Downloads • Specify the following details: –...
  • Page 164: Maintenance

    ‣ only use the associated Endress+Hauser housing seals. 1. Replace defect seals only with original seals from Endress+Hauser. 2. The housing seals must be clean and undamaged when inserted into their grooves. 3. Dry, clean or replace the seals if necessary.
  • Page 165: Measuring And Test Equipment

    Maintenance 13.2 Measuring and test equipment Endress+Hauser offers a wide variety of measuring and test equipment, such as W@M or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services. List of some of the measuring and testing equipment: 13.3...
  • Page 166: Repairs

    • The measuring devices have a modular design. • Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by appropriately trained customers. • Certified devices can only be converted to other certified devices by Endress+Hauser Service or at the factory.
  • Page 167: Endress+Hauser Services

    14.3 Endress+Hauser services Endress+Hauser offers a wide range of services. Your Endress+Hauser Sales Center can provide detailed information on the services. 14.4 Return The measuring device must be returned if it is need of repair or a factory calibration, or if the wrong measuring device has been delivered or ordered.
  • Page 168: Disposing Of The Measuring Device

    Repairs Proline Prowirl R 200 HART 2. Carry out the mounting and connection steps from the "Mounting the measuring device" and "Connecting the measuring device" sections in reverse order. Observe the safety instructions. 14.5.2 Disposing of the measuring device WARNING Danger to personnel and environment from fluids that are hazardous to health.
  • Page 169: Accessories

    Various accessories, which can be ordered with the device or subsequently from Endress +Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 170: For The Sensor

    For intrinsically safe HART communication with FieldCare via the USB interface. HART  Technical Information TI00404F Commubox FXA291 Connects Endress+Hauser field devices with a CDI interface (= Endress+Hauser Common Data Interface) and the USB port of a computer or laptop.  Technical Information TI405C/07 HART Loop Converter...
  • Page 171: Service-Specific Accessories

    Accessories Description Applicator Software for selecting and sizing Endress+Hauser measuring devices: • Choice of measuring devices for industrial requirements • Calculation of all the necessary data for identifying the optimum flowmeter: e.g. nominal diameter, pressure loss, flow velocity and accuracy.
  • Page 172: System Components

    Accessories Proline Prowirl R 200 HART 15.4 System components Accessories Description Memograph M graphic The Memograph M graphic data manager provides information on all the relevant data manager measured variables. Measured values are recorded correctly, limit values are monitored and measuring points analyzed. The data are stored in the 256 MB internal memory and also on a SD card or USB stick.
  • Page 173: Technical Data

    Proline Prowirl R 200 HART Technical data Technical data 16.1 Application Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media. To ensure that the device remains in proper operating condition for its service life, use the measuring device only for media against which the process-wetted materials are sufficiently resistant.
  • Page 174 Technical data Proline Prowirl R 200 HART Calculated measured variables Order code for "Sensor version; DSC sensor; measuring tube" Option Description Measured variable Volume; 316L; 316L Under constant process conditions: • Mass flow Volume; Alloy C22; 316L • Corrected volume flow Volume high-temperature;...
  • Page 175 Proline Prowirl R 200 HART Technical data Liquids Gas/steam [in] [ft³/min] [ft³/min] 3R, 4S 0.77 to 36 3.8 to 480 4R, 6S 1.7 to 81 8.6 to 1 100 6R, 8S 3 to 140 15 to 1 900 8R, 10S 6.8 to 320...
  • Page 176 Technical data Proline Prowirl R 200 HART Reynolds number Flow Internal diameter of measuring tube (corresponds to dimension K) µ Dynamic viscosity ρ Density The Reynolds number, 5 000 together with the density and viscosity of the fluid and the nominal diameter, is used to calculate the corresponding flow rate.
  • Page 177 Proline Prowirl R 200 HART Technical data Steam quality Vibration [m/s] · · D [m] π ² AmpMin [m /h] = ³ · 3600 [s/h] AmpMin ρ [kg/m ] ³ 4 · 1 [kg/m ] ³ [ /s] · · D [ft] π...
  • Page 178 Technical data Proline Prowirl R 200 HART 350 [m/s] · · D [m] π ² [m /h] = ³ · 3600 [s/h] AmpMax ρ [kg/m ] ³ 4 · 1 [kg/m ] ³ 1148 [ /s] · · D [ft] π...
  • Page 179 • Operating pressure to increase accuracy (Endress+Hauser recommends the use of a pressure measuring device for absolute pressure, e.g. Cerabar M or Cerabar S) • Medium temperature to increase accuracy (e.g. iTEMP) •...
  • Page 180: Output

    Technical data Proline Prowirl R 200 HART If the measuring device does not have pressure or temperature compensation , it is recommended that external pressure measurement values be read in so that the following measured variables can be calculated: • Energy flow •...
  • Page 181 Proline Prowirl R 200 HART Technical data Pulse/frequency/switch output Function Can be set to pulse, frequency or switch output Version Passive, open collector Maximum input values • DC 35 V • 50 mA Voltage drop • For ≤ 2 mA: 2 V •...
  • Page 182 Technical data Proline Prowirl R 200 HART Signal on alarm Depending on the interface, failure information is displayed as follows: Current output 4 to 20 mA 4 to 20 mA Failure mode Choose from: • 4 to 20 mA in accordance with NAMUR recommendation NE 43 •...
  • Page 183: Power Supply

    Proline Prowirl R 200 HART Technical data Protocol-specific data Manufacturer ID 0x11 Device type ID 0x0038 HART protocol revision Device description files Information and files under: (DTM, DD) www.endress.com HART load • Min. 250 Ω • Max. 500 Ω System integration For information on system integration, see .→...
  • Page 184 Technical data Proline Prowirl R 200 HART Increase in minimum Order code for "Sensor version; DSC sensor; measuring tube" terminal voltage Option DA: + DC 1 V Mass steam; 316L; 316L (integrated pressure/temperature measurement) Option DB: Mass gas/liquid; 316L; 316L (integrated pressure/temperature...
  • Page 185: Performance Characteristics

    Proline Prowirl R 200 HART Technical data Overvoltage protection The device can be ordered with integrated overvoltage protection for diverse approvals: Order code for "Accessory mounted", option NA "Overvoltage protection" Input voltage range Values correspond to supply voltage specifications →  183 Resistance per channel 2 ⋅...
  • Page 186 Technical data Proline Prowirl R 200 HART Reynolds number 5 000 10 000 Reynolds number for minimum permitted volume flow in measuring tube • Standard • Option N "0.65% volume PremiumCal 5-point, extended turndown" [m/s] · · D [m] π...
  • Page 187 Proline Prowirl R 200 HART Technical data Order code for "Pressure component" Nominal value Pressure ranges and measured errors [bar abs.] Pressure range Maximum measured error [bar abs.] Option D 0.01 ≤ p ≤ 2 0.5 % of 2 bar abs.
  • Page 188 Order code for "Calibration flow", option N "0.65% volume PremiumCal 5-point, extended turndown" Mass flow (user-specific liquids) To specify the system accuracy, Endress+Hauser requires information about the type of liquid and its operating temperature or information in table form about the dependency between the liquid density and the temperature.
  • Page 189: Installation

    Proline Prowirl R 200 HART Technical data 0.40 0.35 0.30 0.25 0.20 0.15 0.10 0.05 V / D ³ 1000 10000 100000 A0034414  31 Repeatability = 0.1 % o.r. at a measured volume [m ] of V = 1000 ⋅ D The repeatability can be improved if the measured volume is increased.
  • Page 190 Technical data Proline Prowirl R 200 HART Temperature tables Observe the interdependencies between the permitted ambient and fluid temperatures when operating the device in hazardous areas. For detailed information on the temperature tables, see the separate document entitled "Safety Instructions" (XA) for the device.
  • Page 191: Process

    Proline Prowirl R 200 HART Technical data Shock resistance Shock, half-sine according to IEC 60068-2-27 • Order code for "Housing", option C "GT20 two-chamber, aluminum, coated, compact", J "GT20 two-chamber, aluminum, coated, remote", K "GT18 two-chamber, 316L, remote" 6 ms, 50 g •...
  • Page 192 Technical data Proline Prowirl R 200 HART Seals Order code for "DSC sensor seal" Option Description Medium temperature range Graphite (standard) –200 to +400 °C (–328 to +752 °F) Viton –15 to +175 °C (+5 to +347 °F) Gylon –200 to +260 °C (–328 to +500 °F) Kalrez –20 to +275 °C (–4 to +527 °F)
  • Page 193: 16.10 Mechanical Construction

    Proline Prowirl R 200 HART Technical data Sensor Maximum sensor measuring range Lower (LRL) Upper (URL) [bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)] 2 bar (30 psi) 0 (0) +2 (+30) 6.7 (100.5) 10 (150) 4 bar (60 psi)
  • Page 194 Technical data Proline Prowirl R 200 HART Internal diameter Weight [lbs] [in] [in] Order code for "Housing", option C Order code for "Housing", option B "GT20 two-chamber, aluminum, coated, "GT18 two-chamber, 316L, compact" compact" ½ 18.0 23.9 1½R 22.4 28.3 1½...
  • Page 195 Proline Prowirl R 200 HART Technical data Weight in US units All values (weight) refer to devices with ASME B16.5, Class 300/Sch. 40 flanges. Weight information in [lbs]. Internal diameter Weight [lbs] [in] [in] sensor connection housing sensor connection housing Order code for "Housing", option J...
  • Page 196 Technical data Proline Prowirl R 200 HART Pressure rating Weight [mm] [kg] Class 150 Class 300 Class 150 Class 300 Class 150 Class 300 Class 150 12.3 Class 300 15.8 Class 150 25.7 Class 300 27.5 ASME Pressure rating Weight...
  • Page 197 Proline Prowirl R 200 HART Technical data Pressure rating Weight [in] [lbs] Class 150 27.0 Class 300 35.0 Class 150 57.0 Class 300 61.0 ASME Materials Transmitter housing Compact version • Order code for "Housing", option B "GT18 two-chamber, 316L, compact": Stainless steel, CF3M •...
  • Page 198 Technical data Proline Prowirl R 200 HART Order code for "Housing": option C "GT20 two-chamber, aluminum, coated, compact", option J "GT20, two-chamber, aluminum, coated remote" Applies also to the following device versions in combination with HART communication mode: Order code for "Sensor version; DSC sensor; measuring tube", option DA "mass steam;...
  • Page 199 Proline Prowirl R 200 HART Technical data Pressure ratings PN 10/16/25/40, Class 150/300, as well as JIS 10K/20K: Parts in contact with medium (marked as "wet" on the DSC sensor flange): • Stainless steel 1.4404 and 316 and 316L • Compliant with: –...
  • Page 200 Technical data Proline Prowirl R 200 HART The following materials are available depending on the pressure rating: Stainless steel, multiple certifications, 1.4404/F316/F316L) List of all available process connections Seals • Graphite (standard) Sigraflex foil (BAM-tested for oxygen applications, "high-grade in the context of TA- Luft Clean Air Guidelines")
  • Page 201: 16.11 Operability

    Proline Prowirl R 200 HART Technical data 16.11 Operability Languages Can be operated in the following languages: • Via local display: English, German, French, Spanish, Italian, Dutch, Portuguese, Polish, Russian, Swedish, Turkish, Chinese, Japanese, Korean, Bahasa (Indonesian), Vietnamese, Czech • Via "FieldCare" operating tool:...
  • Page 202: 16.12 Certificates And Approvals

    EU Directives. These are listed in the corresponding EU Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications and Media Authority (ACMA)".
  • Page 203: 16.13 Application Packages

    The application packages can be ordered with the device or subsequently from Endress+Hauser. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 204: 16.14 Accessories

    • The W@M Device Viewer : Enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate. Standard documentation...
  • Page 205 Proline Prowirl R 200 HART Technical data Contents Documentation code EAC Ex d XA01684D EAC Ex nA XA01685D Special documentation Contents Documentation code Information on the Pressure Equipment Directive SD01614D Functional Safety Manual SD02025D Contents Documentation code HART FOUNDATION Fieldbus...
  • Page 206: Index

    Index Proline Prowirl R 200 HART Index Access authorization to parameters Declaration of Conformity ..... 10 Read access ......63 Define access code .
  • Page 207 Proline Prowirl R 200 HART Index Field Communicator 475 ....64 Field Communicator ..... . 68 Field Xpert SFX350/SFX370 .
  • Page 208 Index Proline Prowirl R 200 HART Manufacturer ID ......69 Operation ....... . 136 Manufacturing date .
  • Page 209 Proline Prowirl R 200 HART Index Protecting parameter settings ....124 Showing data logging ..... . . 143 Signal on alarm .
  • Page 210 Index Proline Prowirl R 200 HART System design Wizard Measuring system ..... . . 173 Current input ......82 see Measuring device design Current output 1 to n .
  • Page 212 www.addresses.endress.com...

Table of Contents