Setup And Installation; Installation Site - Eizo CuratOR LX491W Instructions For Use Manual

Full hd 49" lcd monitor
Table of Contents

Advertisement

Setup and installation

4.1 Installation site

4 Setup and installation
CAUTION
Changes to device
Do not make any mechanical or electric changes to the device.
EIZO GmbH is not liable for changes made to the device.
4.1 Installation site
NOTICE
The power switch and connections must be accessible at all times
When installing and connecting the monitor, ensure that the power switch and the connec-
tions are accessible at all times.
NOTICE
Condensation
If the device is brought into a warm environment from a cold one, condensation may form
in the device. This could result in a short circuit when switching on the device, damaging it.
• Wait until the condensed water has evaporated, including that inside the device, before
you switch it on. This can take several hours.
NOTICE
Overheating
Ventilation holes are located on the rear of the housing.
If the ventilation holes are covered or closed, the heat generated in the monitor will not be
dissipated sufficiently.
• Do not cover the ventilation holes.
• Do not close the ventilation holes.
• The minimum distance from the back and side of the monitor to the wall must be 10 cm,
and at least 15 cm from other devices.
• The ambient temperature must be in the acceptable range of +5 °C to +40 °C.
NOTICE
Dusty environment
The monitor is intended for use in the clean environment of medical diagnostics. In dusty
environments, ventilation holes in the back can allow dust to penetrate into the monitor.
In the worst case, deposits are possible which become evident as dark spots in a white
picture and result in deterioration of the luminance.
• Protect the monitor from dust, for example through appropriate construction measures
at the installation site.
14
CuratOR LX491W
Instructions for Use, 02/2018

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents