Setup and installation

CAUTION
Changes to device
Do not make any mechanical or electric changes to the device.
EIZO GmbH will not be held liable if changes are made to the device.
4.1

Installation site

Note the following conditions at the installation location.
NOTICE
The power switch and connections must be accessible at all times
When installing and connecting the monitor, ensure that the power switch and the connec‐
tions are accessible at all times.
NOTICE
Condensation
If the device is brought into a warm environment from a cold one, condensation may form
in the device. This could result in a short circuit when switching on the device, damaging it.
• Wait until the condensed water has evaporated, including that inside the device, before
you switch it on. This can take several hours.
Sufficient ventilation
Ventilation holes are located on the rear of the housing.
If the ventilation holes are covered or closed, the heat generated in the monitor will not be
dissipated sufficiently.
Do not cover the ventilation holes.
Do not close the ventilation holes.
The minimum distance from the back and side of the monitor to the wall must be 10 cm,
and at least 15 cm from other devices.
Avoid dusty environments
The monitor is intended for use in the clean environment of medical diagnostics. In dusty
environments, ventilation holes in the back can allow dust to penetrate into the monitor. In
the worst case, deposits are possible which become evident as dark spots in a white picture
and result in deterioration of the luminance.
Protect the monitor from dust, for example through appropriate building measures at the
installation location.
During transport, use the original packaging or service packaging.
RadiForce LX600W
Instruction for Use, 11/2013
Setup and installation
4.1 Installation site
15