Autotest Results For Far End Source Mode - Fluke CertiFiber Pro User Manual

Optical loss test set
Hide thumbs Also See for CertiFiber Pro:
Table of Contents

Advertisement

CertiFiber Pro Fiber Optical Loss Test Set
Users Manual
3-1
On the SET REFERENCE, when the set reference or TRC
verification procedure is completed, tap NEXT to see how to
connect to the fiber under test.
3-2
Clean and inspect the all connectors.
3-3
Make the connections to do the test on the fiber, as shown on
the screen and in Figure 24, then tap HOME.
3-4
Tap TEST on the main tester or press  on the main or
remote tester.
3-5
If the CHECK FIBER CONNECTIONS screen shows an open
fiber:
Make sure that all connections are good and no fibers
have damage. Use the VFL to make sure the fiber under
test has continuity.
If you are connected to a fiber at a patch panel, and you
are not sure you are connected to the correct fiber,
connect the main tester's INPUT fiber to different
connections until the test continues.
3-6
If Auto Save is on, the tester uses the next ID to save the
results.
If Auto Save is off, the SAVE RESULT screen shows the next
ID available. You can change the ID if necessary.

Autotest Results for Far End Source Mode

Figure 25 shows an example of Autotest results for Far End Source
mode.
The Autotest in Far End Source mode does not
show a PASS/FAIL result, limit, or margin if the
selected test limit uses fiber length to calculate
loss. An example of such a limit is the TIA-568C
Fiber Backbone limit. The tester does not measure
length in Far End Source mode.
78