Siemens SITRANS Probe LU Operating Instructions Manual page 55

Ultraschall transmitter
Hide thumbs Also See for SITRANS Probe LU:
Table of Contents

Advertisement

P650 Offset Calibration
Calibrates Empty (P006) if the reported level is consistently high or low by a fixed amount
(stored in P652).
Values
Related
Before using this feature, verify the following parameters are correct:
Empty (P006)
Temperature (P664)
Offset Calibration
Begin with a steady level.
1.
Press the measurement key
2.
Repeat step 1 at least five times, to verify repeatability.
3.
Measure the actual reading (use tape measure).
4.
Key in the actual value, and press ENTER
The deviation between the entered Empty (P006) value and the calibrated Empty value is
stored in Offset Correction (P652).
P651 Sound Velocity Calibration
Calibrates the speed of sound constant.
Values
Related
Use P651 under the following conditions:
The acoustic beam atmosphere is other than air.
The acoustic beam atmosphere temperature is unknown.
The reading accuracy is acceptable only at higher material levels.
For best results, calibrate with the level at a known value close to empty.
A5E32337695
SITRANS Probe LU (HART) – OPERATING INSTRUCTIONS
Range
–99999 to 99999
P006 Empty
P652 Offset Correction
P664 Temperature
to display the calculated reading.
Range
–99999 to 999999
P653 Velocity
o
P654 Velocity at 20
C
.
Page 49