Loop Voltage Versus Loop Resistance; Is Safety Barrier Selection; How To Select A Passive Barrier For Sitrans Probe Lu - Siemens SITRANS Probe LU Operating Instructions Manual

Ultraschall transmitter
Hide thumbs Also See for SITRANS Probe LU:
Table of Contents

Advertisement

Loop Voltage versus Loop Resistance

1000
900
800
700
600
500
400
300
200
100
0
10.00

IS Safety Barrier Selection

Selecting a suitable barrier or power supply requires knowledge about Intrinsic Safety
and the application. It is the responsibility of the installer to ensure that the intrinsically
safe installation complies with both the apparatus approval requirements and the
relevant national code of practice.

How to select a passive barrier for SITRANS Probe LU

1.
Make sure that the barrier safety description is suitable for the SITRANS Probe LU
Intrinsically Safe (IS) input parameters.
2.
Determine the maximum end-to-end resistance of the barrier (Re-e) from the data
sheet.
3.
Determine other loop resistance (Rloop): for example, sense resistance, displays,
and/or PLC inputs.
4.
Calculate Rworking = Re-e + Rloop.
A5E32337695
SITRANS Probe LU (HART) – OPERATING INSTRUCTIONS
Power Supply Requirements
(17.8v, 300 ohms)
RL=272.7*VL - 4554.1
(16.7v, 0 ohms)
15.00
Loop Voltage-V
(Source voltage)
(30.0v,845 ohms)
RL=44.6*VL - 493.4
ALLOWABLE OPERATING
AREA
20.00
25.00
L
30.00
Page 95

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents